Upon power on the counter module defaults to the following setup:
Baud rate: 9600 baud, 8 data bits, no parity, 1 stop bit
Cycle rate: 10 Hz
Analog channels: Channel 0 (Larmor signal level) enabled, channels 1-5 disabled
Julian Clock: Disabled
Output Format: ASCII
The default output data stream contains all printable ASCII characters with each sample
terminated with a carriage return/ line feed sequence. Table 5 illustrates an example of this
format.
Character
#
Description
1
An ASCII '$' (marks first character of data stream)
2
an ASCII '1' or a blank (depending on whether Mag reading is above or below
99999.999nT).
3-7
5 digits of Mag data
8
an ASCII decimal point ['.']
9-11
3 more digits of Mag data
12
an ASCII comma [',']
13-16
4 digits of A/D channel 0 (9999 full scale, 0 to +5 volts in). This channel is internal and
contains the signal level of the magnetometer.
17
an ASCII carriage return
18
an ASCII line feed
If the data were captured to a file and then copied to a line printer the printout would look
something like this:
$ 99778.131,3749
$ 99890.376,3687
$ 99955.517,3545
$ 99998.293,3472
$100078.835,3329
$100032.071,3381
$ 99979.159,3498
$ 86778.508,3514
$ 78778.216,3645
Table 5. Example of CM-201 default output data stream.