Check & Adjust
4 Check & Adjust
4.1 Overview
Description
GeoMax instruments are manufactured, assembled and adjusted to a high quality.
Quick temperature changes, shock or stress can cause deviations and decrease
the instrument accuracy. It is therefore recommended to calibrate the instrument
from time to time. This can be done in the field by running through specific meas-
urement procedures. The procedures are guided and have to be followed carefully
and precisely as described in the following chapters. Some other instrument
errors and mechanical parts can be adjusted mechanically.
Electronic
adjustment
The following instrument errors can be checked and adjusted electronically:
Every angle measured in the daily work is corrected automatically if the
compensator and the Hz-corrections are activated in the instrument configu-
ration. Select Main Menu: Config...
\
Instrument Settings...
\
Compensator
to check the settings.
l, t
Compensator longitudinal and transversal index errors
i
Vertical index error, related to the standing axis
c
Horizontal collimation error, also called line of sight error
a
Tilting axis error
Aim360
Aim360 zero point error for Hz and V - option
Содержание Zoom80 C
Страница 1: ...GeoMax Zoom80 Series User Manual Version 1 0 ...
Страница 25: ...Description of the System Zoom80 25 Instrument components for RCS a ZRT80 b Bluetooth port a b ...
Страница 139: ...Safety Directions Zoom80 139 a Laser beam b Exit for laser beam b Zoom80_016 a ...
Страница 184: ...Zoom80 184 Technical Data Antenna Type 2 dipole antenna Gain 2 dBi Connector Special customized SMB ...
Страница 193: ...Index Zoom80 193 ...