RSS-2-300W User Manual
34
RSS-2-300W Non-Contact Surface Velocity Radar
5. Make sure that the COM port is not already open
Only one application may use a single COM port in the system. Make sure that
no other open application uses the same COM port that you are trying to open.
6. Try restarting the application and/or the computer
Close and reopen the configurator application and try to establish the
connection again. Restart your computer and try to establish the connection to
the instrument again.
The instrument
does not respond
over Modbus (RS-
485) interface.
1. Make sure that the cable is properly connected to the instrument
Check that the cable M12 circular connector is firmly connected and screwed
to the instrument. It is not sufficient to simply attach the cable connector to the
instrument connector, the cable connector must be screwed into the instru-
ment connector.
2. Check power supply
Make sure that the power is being supplied to the instrument. A direct-current
voltage, between 9 and 27 Volts must be connected to brown (+) and white (-)
wires of the instrument cable. The power supply must be able to deliver at least
500 mA of current.
3. Check RS-485 connector
Make sure that the RS-485 lines on the instrument cable are properly
connected to the RS-485 connector. The dark red wire should be connected to
D+ line, and the orange line should be connected to D- line.
Make sure that the D+ and D- lines are not swapped.
Make sure that you have correctly identified and connected the dark red
(magenta) wire, as there is also another bright red wire that is used to provide
output power for 4-20mA output. It is possible to misidentify the wire and to
connect the bright red wire to D+ instead of dark red wire.
4. Check that you are using the correct slave device ID
The default Modbus device ID is 1. The Modbus device ID can be changed by
connecting the instrument to the PC application (over RS-232 connection).
Connect the instrument to the PC using RS-232 connection. Open the Geolux
Instrument Configurator application and establish a connection between the
instrument and the PC. Then check the
Modbus ID
parameter and make sure
that it is the same as the slave device ID used in issued Modbus requests.
5. Make sure that there are no two devices on the bus with the same Device ID.
Modbus allows to have multiple devices connected on the same bus
simultaneously. Each device must have a unique slave device ID assigned,
so that the bus master can distinguish between the devices. If two or more
devices are assigned the same slave device ID, a bus conflict will happen and
prohibit the master to correctly communicate with the slave devices. To resolve
this problem, change the instrument’s slave device ID to a unique number
through the Geolux Instrument Configurator PC application.