
14
5. TROUBLESHOOTING
Maintenance and troubleshooting of rebar strain meters is confined to periodic checks of cable
connections and maintenance of terminals. Once installed, the strain meters are usually
inaccessible and remedial action is limited.
Gauges should not be opened in the field.
Should difficulties arise, consult the following list of problems and possible solutions. Return
any faulty gauges to the factory. For additional troubleshooting and support contact Geokon.
Symptom: Thermistor resistance is too high
It is likely that there is an open circuit. Check all connections, terminals, and plugs. If a cut is
located in the cable, splice according to instructions in Section 2.5.
Symptom: Thermistor resistance is too low
It is likely that there is a short. Check all connections, terminals, and plugs. If a short is
located in the cable, splice according to instructions in Section 2.5.
Water may have penetrated the interior of the strain meter. There is no remedial action.
Symptom: Strain Meter Readings are Unstable
Is the readout box position set correctly? If using a datalogger to record readings
automatically, are the swept frequency excitation settings correct?
Is there a source of electrical noise nearby? Likely candidates are generators, motors, arc
welding equipment, high voltage lines, etc. If possible, move the instrument cable away from
power lines and electrical equipment or install electronic filtering.
Make sure the shield drain wire is connected to ground whether using a portable readout or
datalogger. Connect the shield drain wire to the readout using the blue clip. (Green for the
GK-401.)
Does the readout work with another strain meter? If not, it may have a low battery or
possibly be malfunctioning.
Symptom: Strain Meter Fails to Read
Is the cable cut or crushed? Check the resistance of the cable by connecting an ohmmeter to
the gauge leads. Table 2 shows the expected resistance for the various wire combinations;
Table 3 is provided for the user to fill in the actual resistance found. Cable resistance is
approximately 14.7
Ω
per 1,000 feet (48.5
Ω
per km); multiply this factor by two to account
for both directions. If the resistance is very high or infinite (megohms), the cable is probably
broken or cut. If the resistance is very low (
<
20
Ω
), the gauge conductors may be shorted. If a
cut or a short is located in the cable, splice according to the instructions in Section 2.5.
Does the readout or datalogger work with another gauge? If not, it may have a low battery or
possibly be malfunctioning.
Содержание 4911
Страница 2: ......
Страница 4: ......
Страница 9: ...3 Figure 3 Model 4911 Sister Bar Installation Figure 4 Model 4911 Sister Bar Installation Detail ...
Страница 27: ...21 APPENDIX D SAMPLE CALIBRATION REPORT Figure 10 Sample Model 4911 Calibration Report ...