3. TAKING READINGS
3.1 GK-404 Readout Box
The Model GK-404 Vibrating Wire Readout is a portable, low-power, handheld unit that is
capable of running for more than 20 hours continuously on two AA batteries. It is designed for
the readout of all Geokon vibrating wire gages and transducers, and is capable of displaying the
reading in either digits, frequency (Hz), period (µs), or microstrain (µε). The GK-404 also
displays the temperature of the load cell (embedded thermistor) with a resolution of 0.1 °C.
3.1.1 Operating the GK-404
Before use, attach the flying leads to the GK-404 by aligning the red circle on the silver
“Lemo” connector of the flying leads with the red line on the top of the GK-404 (Figure
16). Insert the Lemo connector into the GK-404 until it locks into place.
Figure 16 - Lemo Connector to GK-404
Connect each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
To turn the GK-404 on, press the “ON/OFF” button on the front panel of the unit. The
initial startup screen will be displayed. After approximately one second, the GK-404 will
start taking readings and display them based on the settings of the POS and MODE
buttons.
The unit display (from left to right) is as follows:
•
The current Position: Set by the
POS
button. Displayed as a letter A through F.
•
The current Reading: Set by the
MODE
button. Displayed as a numeric value
followed by the unit of measure.
•
Temperature reading of the attached gage in degrees Celsius.
Use the
POS
button to select position
B
and the
MODE
button to select
Dg
(digits).
(Other functions can be selected as described in the GK-404 Manual.)
The GK-404 will continue to take measurements and display readings until the unit is
turned off, either manually, or if enabled, by the Auto-Off timer. If the no reading
displays or the reading is unstable, see Section 5 for troubleshooting suggestions.
For further information, please see the GK-404 manual.