![Geokon 4500 Series Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/geokon/4500-series/4500-series_instruction-manual_2214559039.webp)
MODEL 4500 SERIES VIBRATING WIRE PIEZOMETER
| PIEZOMETER PRESSURE AND WATER LEVEL |
29
APPENDIX F.
PIEZOMETER PRESSURE AND WATER
LEVEL
Frequently, when using a dip meter to check water levels in an open well, it
happens that the water level, computed from a piezometer reading taken at the
bottom of the well, does not agree with the water level measured directly by the
dip meter. This will happen when the specific gravity of the water is not 1 gm/cc,
the water is brackish or muddy, or both. It will also occur when there is a flow of
water up or down the borehole. In addition, if the piezometer is removed to
make room for the dip meter, the volume of the piezometer and cable displaces
an equal volume of water, which can cause changes of the water level in a small
diameter well.