12
EN
FACTORY SETTINGS
Minimum reading
Quadrant signal tone
Unit of angle measurement
Automatic shut-off
Vertical angle measuring mode
Automatic compensation
Data transmission
1“
OFF
GON
OFF
ZENITH ANGLE
OFF
OFF
SETTTING METHODS
Power on the instrument.
Press „FUNC“ and then the ON/OFF
key.
Press the buttons and to select the required values (1 -
7).
The parameters set will be available until the next parameter
setting - even if the instrument is powered off.
Δ
Δ
Numbers 1 to 7
ERROR DISPLAY
DISPLAY
B
E00
E01
E02
E03
E04
E05
E06
ERROR DESCRIPTION AND DISPOSAL
The vertical compensator is out of the compensating range. Level the instrument
again.
The collimate rotation is too fast. Press the 0SET key to set 0. If “E00” is
shown again the instrument needs to be repaired.
The telescope turning is too fast. Press the V/% key to set 0 for the vertical plate
specification.
Interior error of the horizontal angle measuring system. Power on the instrument
again. If the error is shown again the instrument needs to be repaired.
Interior error of the vertical angle measuring system. If the error is shown again
the instrument needs to be repaired.
Difference between every value exceeding 30“ during the angle repeated mea-
surement. Press the 0SET key and measure repeatedly.
The number of angle repeated measurement exceeds 15. Press key and measure
repeatedly.
Errors during the process of vertical angle 0set or adjusting 0set when the clamp
angle to the horizon exceeds 45°. Adjust repeatedly.