U-LINK User Manual 

Revision 1.0 

 

 

23 

 

Command:  *GWL 

Answer: 

PWC: 1064<CR><LF> 

 

4.6.4.  Control 

*ANT - Set Anticipation  

 
This command is used to enable or disable the anticipation processing when the device is reading from a 
wattmeter.  The anticipation is a software-based acceleration algorithm that provides faster readings using 
the detector’s calibration. 
 

Command 

Parameters 

Answer 

ANT 

1: On 
0: Off 

None 

 

Default: 

 On 

 

  Example 

 

The following example sets the anticipation On. 

 

Command:  *ANT1 

Answer: 

 

*GAN - Get Anticipation Status 

 
This command returns the anticipation status. If the anticipation is not available, it will always be at “off”. 
 

Command 

Parameters 

Answer 

GAN 

None 

1: On 
0: Off 

 

  Example 

 

Command:  *GAN 

Answer: 

Anticipation: 0<CR><LF> 

 

*AVG - Noise Suppression Averaging 

 
Sets the sampling size of the noise suppression. 

For pyroelectric detectors and UM detectors only

 
The  joulemeter  mode  has  a  special  proprietary  algorithm  that  can  lower  the  noise-induced  error  when 
reading low energy levels, or energy readings of any level with noise present. This feature greatly reduces 
the effect of noise on the peak-to-peak measurement in joulemeter mode. The monitor will need to measure 
the number of pulses selected in the Sampling Size before the algorithm will settle to the noise suppressed 
value. Once the readings have stabilized, any subsequent reading will be stable until the Sampling Size is 
changed. The system will then stabilize to the new value. Larger sampling sizes will result in more noise 
suppression. Noise suppression works best with the external trigger. This function will greatly improve the 
accuracy of the lowest scale or in any scale when used in a noisy environment. 
 
The parameter must consist of exactly 3 digits. 
 

Command 

Parameters 

Answer 

Содержание U-LINK

Страница 1: ...U LINK User Manual Revision 1 0 i WARRANTY...

Страница 2: ...EO office to obtain a return authorization number Return the material to the address below Contacting Gentec Electro Optics Inc Gentec Electro Optics Inc 445 St Jean Baptiste Suite 160 Qu bec QC Canad...

Страница 3: ...les These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate rad...

Страница 4: ...3 User Interface 9 4 Serial Communication 9 4 1 Description 9 4 2 Setting up Communication to the U LINK 9 4 2 1 Verify the COM Port 9 4 2 2 Connect the U LINK 9 4 3 Serial Command Format 10 4 4 Bina...

Страница 5: ...onnector on RS 232 U LINK Cable 6 Figure 6 Opening U LINK for Recycling Disposal 37 Table of Tables Table 1 Table of Specifications 2 Table 2 Multi Function Port Feature Availability 5 Table 3 RS 232...

Страница 6: ...or RS 232 port to a personal computer The U LINK USB and RS 232 versions have enhanced communication capabilities that take advantage of the USB or RS 232 port for data acquisition and remote control...

Страница 7: ...ected scale 3754 energy mode pyroelectric power mode and fast power mode at 100Hz Monitor Accuracy 0 5 3 V from 20 to full scale1 Response Time accelerated 2 Head dependent Accelerated 1 second Data t...

Страница 8: ...minimum trigger pulse width 550 1 549 sec Therefore the acceptable window for the beginning of the external trigger pulse is from 110 sec before the laser pulse begins to 549 sec after the laser pulse...

Страница 9: ...U LINK User Manual Revision 1 0 4 Figure 1 Valid External Trigger Window 1 4 CONNECTIONS Figure 2 U LINK USB Front Panel Figure 3 U LINK RS232 Front Panel...

Страница 10: ...as a chart recorder a computer with an analog interface a voltmeter etc Use this port with Gentec EO Part Number 100 201958 Analog Out Cable for this feature The output signal represents a DC analog v...

Страница 11: ...Gentec EO Part Number 201860 Do not use an RS232 adapter with a U LINK USB Table 3 RS 232 U LINK Cable Pinout Pin Number Signal 1 No Connection 2 TX 3 RX 4 No Connection 5 Signal GND 6 Do Not Connect...

Страница 12: ...based on spectral data measured from that specific detector The U LINK may not recognize some of the earlier heads This DB 15 connector though similar to that of the former TPM 310 and TPM 330 monitor...

Страница 13: ...lock the connector into place 6 Connect U LINK to your computer via USB 7 Start the PC Gentec EO software Click the Connect button in the top left corner and select your U LINK from the list of devic...

Страница 14: ...nds which must follow rules stated in this section The USB class used by U LINK USB is a CDC or Communications Device Class This means it shows up in the host PC as a COM port but it is not a COM port...

Страница 15: ...cute the following sequence File Properties Settings ASCII setup Character delay add a few milliseconds of delay OK When you end the session HyperTerminal asks if you want to save your settings To avo...

Страница 16: ...s this is the High byte and X is the upper binary data If the measurement is over range this byte is 0xFE Low byte 1XXX XXYY in binary 1 indicates this is the Low byte X is the lower binary data and Y...

Страница 17: ...ata bytes are 0xA0B6 If these bytes are 0xFE7F an out of scale condition exists If no out of scale condition exits mask off bit 7 of each byte then follow steps 3 to 6 of the procedure to decode CAU r...

Страница 18: ...tor MUL Set User Multiplier Sets the multiplier value GUM Get User Multiplier Returns the current multiplier value OFF Set User Offset Sets the offset value GUO Get User Offset Returns the current off...

Страница 19: ...100 picowatts or picojoules 25 3 watts or joules 05 300 picowatts or picojoules 26 10 watts or joules 06 1 nanowatt or nanojoule 27 30 watts or joules 07 3 nanowatts or nanojoules 28 100 watts or joul...

Страница 20: ...er scale Command Parameters Answer SSD None None GCR Get Current Scale Index This command returns the scale index between 0 and 41 Please refer to Set Scale command SCS details for the complete scale...

Страница 21: ...of valid scales The following example is for a UP19K which can have the following scales 100 mW 300 mW 1 W 3 W 10 W 30 W 100 W Example Command DVS Answer 22 100 0 m CR LF 23 300 0 m CR LF 24 1 000 CR...

Страница 22: ...de in W Energy Mode in J or Single Shot Energy Mode in J SSE Command Parameters Answer GMD None POWER 0 ENERGY 1 SSE 2 Example Command GMD Answer Mode 0 CR LF 4 6 2 Data Acquisition CVU Query Current...

Страница 23: ...fer to section 4 4 Command Parameters Answer CAU None Data in ASCII Scientific notation or in binary format Examples For example with a wattmeter a reading around 500 milliwatts would be displayed lik...

Страница 24: ...R LF 5 066012e 01 32 0 CR LF 5 066015e 01 32 0 CR LF 5 066021e 01 32 0 CR LF CTU Send Current Value with Frequency U LINK will send the current measurement and the pulse repetition rate in Hz They are...

Страница 25: ...CR LF GRR Get Laser Frequency This command is used to get the laser frequency This is for joulemeters only Command Parameters Answer GRR None Data in ASCII SS1 Set Binary Joulemeter Mode This command...

Страница 26: ...eters Answer QTM None Window length in seconds Default 0 2 Seconds Example Command QTM Answer 0 2 CR LF 4 6 3 Setup PWC Set Personal Wavelength Correction in nm This command is used to specify the wav...

Страница 27: ...he desired wavelength does not have 5 digits you must enter a zero padded number For example to set the wavelength at 10 6 micron you must enter 010 6 Please note that the resolution is limited as fol...

Страница 28: ...of the noise suppression For pyroelectric detectors and UM detectors only The joulemeter mode has a special proprietary algorithm that can lower the noise induced error when reading low energy levels...

Страница 29: ...nt This command does not work with photodiodes Please refer to SDZ Command Parameters Answer SOU None Autoscale Please Wait Done Fixed scale None Example when in Auto Scale Command SOU Answer Please W...

Страница 30: ...e Fixed scale None Example when in Auto Scale Command SDZ Answer Please Wait CR LF Done CR LF MUL Set User Multiplier This command is used to set the value of the user specified multiplier Command Par...

Страница 31: ...Offset are added to the calculation GUO Get User Offset This command returns the offset value Command Parameters Answer GUO None Current offset value Example Command GUO Answer User Offset 1 5000000E...

Страница 32: ...ollowing example sets the attenuator On this means that the attenuator in on the detector Command ATT1 Answer GAT Get Attenuator This command returns the attenuator status If the attenuator is not ava...

Страница 33: ...100 201956 to synchronize the two monitors Command Parameters Answer SYN 1 On 0 Off None Example Command SYN1 Answer AOB Set Analog Output Scale Watts Joules Sets the output scale of the Analog Outpu...

Страница 34: ...t used for THz Series B V2 heads that require the use of the Lock In Amplifier The value is in seconds and must be entered in the same format as the below example using exactly five characters Command...

Страница 35: ...end the command again at the new baud rate If you want to update the monitor s firmware you will have to change the baud rate to 115200 Command Parameters Answer BPS 0 9600 1 19200 2 38400 3 57600 4 1...

Страница 36: ...000 0 Current scale MSB refer to scale index SCS 0 0008 0019 25 Maximum scale LSB refer to scale index SCS 0 0009 0000 0 Maximum scale MSB refer to scale index SCS 0 000A 0011 17 Minimum scale LSB ref...

Страница 37: ...a valid line while 1 is the end of the structure The next 4 bytes represent the address line and the last 4 bytes are the actual value The values are written on 32 bits which means that all the values...

Страница 38: ...acter 0 0022 0000 The rest of the characters aren t valid until line 002A 0 0023 0000 0 0024 1F 00 0 0025 40 03 0 0026 00 1A 0 0027 0000 0 0028 E1 20 0 0029 00 3A 0 002A 39 31 1 9 Detector name You mu...

Страница 39: ...refer to the M LINK manual for more details Once activated the U LINK will stay in this mode even after being unplugged The M LINK SQL command will not function Command Parameters Answer MLK 1 On 0 Of...

Страница 40: ...55011 2017 Class A CISPR 11Ed6 1 2016 Class A Conducted Emissions Pass EN 61326 1_Ed2 2013 IEC 61326 1_Ed2 2012 Heavy Industrial EN61000 3 2 2014 IEC61000 3 2 2014 Power Line Harmonics Pass EN 61326 1...

Страница 41: ...o normal operation If a range of performance is specified as normal operation the DUT s performance was outside of this range during the application of the test but recovered on its own and operated w...

Страница 42: ...ening the device will void the U LINK s warranty The complete device contains 1 Interface device 1 Power supply RS 232 model power supply not made by Gentec EO 1 USB cable USB model 1 Calibration cert...

Страница 43: ......

Отзывы: