SDX-1239 Operating Instructions
Rev - 3/21/2017
Page 18 of 22
Example.
Send PLY8.20337E-9, -4.99267e-5, 1.09595CRLF
Reply: OK
PRV
Read the sensor responsivity in V/J. Query only.
Example.
Send: prvCRLF
Reply: 12345.00CRLF
Reading SDX-1239 Data
In Measure Mode, the SDX-1239 Sends data at 100ms intervals
. The data is sent in
the format:
Pulse Data, Pulse Period CRLF
Pulse Data
FRDDD
F is the error code.
R is the range index.
DDD are the pulse data counts with 3072 being full scale.
Period Data
PPPPPPPP
PPPPPPPP is the value of the Period Counter. To convert it into the period, divide it by
135E6.