FirstResponder

®

 Sterilizer Manual 

 

VKM180403

 

Genlantis

 

Page 2 of 5 

6044 Cornerstone Court W., Ste. E.  San Diego, CA 92121  

  Toll Free (888) 428-0558  

  U.S. & Canada (858) 457-1919 

 

www.genlantis.com

 

2018 Gene Therapy Systems, Inc. (DBA: Genlantis). 

I.

 

Positioning the Unit 

The FirstResponder

®

 Sterilizer is portable and ready to plug in and use directly 

out of the box.  Its effectiveness depends on how the generated is dispersed 
through ambient air.  We recommend that users place the unit as centrally as 
possible in the vehicle or room, or closest to known or suspected heavy 
contamination areas. 

 

II.

 

Preparing the Area to Be Sterilized 

1.

 

Clean  area  or  cabinet  that  needs  sterilization  from  heavy  or  excess 
contaminants using wipes or cleaning solutions like soap, ethanol, bleach, 
etc.  

 

2.

 

Seal room or vehicle and verify that there are no gaps or openings.  An 
airtight space will prevent TO gas from escaping and will allow the buildup 
of TO molecules.  If any spaces cannot be closed off, it is necessary to 
seal openings with plastic sheeting and/or tape. 

3.

 

 

4.

 

Use  a  copy  of  the  “CAUTION”  page  in  this  manual (Page  7),  or  make  your  own  note  and  prepare  to  affix  on  the  outside  of  the  first 
responder vehicle or room door indicating that a sterilization cycle is in progress.  This is necessary to avoid disrupting a sterilization cycle 
and to eliminate accidental human exposure to high levels of TO. 

 

III.

 

Operating the Unit 

 

General Notes 

a.

 

Different  materials  will  react  differently  to  TO.  Consult  Appendix  A  on  page  4  for 
guidance. 

b.

 

For safety reasons, the unit will not operate if the lid is open. 

c.

 

The unit can be set to operate between 00:01 and 99:99 minutes for either of the TO 
or scrub cycles. 

d.

 

We recommend users keep a log of hours used; this will help with knowing when to 
replace the Scrub Filter (recommended after each 3,000 hours). 

 

To set the run time for the 

ozone cycle

For recommended sterilization times based on size of area, see Table 1 below. 

 

a.

 

Press the [OZONE TIME] button; the digital display will light up and read either “

0000

” or a previously set time. 

b.

 

Press [TIME SELECT] button; the left-most digit will blink; 

c.

 

Press [SET TIME] up or down arrow button repeatedly until desired time is selected. 

d.

 

Press [TIME SELECT] button to move to next digit, repeat Step c, then press [SET TIME] to move to the next digit. 

e.

 

Repeat Step d. one last time to set the last digit; the display now should read the ozone time with no blinking digits. 

 

To set the run time for the 

scrub cycle

For recommended scrubbing times based on size of area, see Table 1 below. 

 

f.

 

Press the [SCRUB TIME] button; the digital display will read either “

0000

” or a previously set time. 

g.

 

Press [TIME SELECT] button; the left-most digit will blink; 

h.

 

Press [SET TIME] up or down arrow button repeatedly until desired time is selected. 

i.

 

Press [TIME SELECT] button to move to next digit, repeat Step c, then press [SET TIME] to move to the next digit. 

j.

 

Repeat Step d. one last time to set the last digit; the display now should read the ozone time with no blinking digits. 

k.

 

If needed or desired, set a lab timer with the total sterilization and scrubbing cycle times as a reminder before you run the unit.  Once 
the ozone and scrub cycles have been set, press the [RUN] button; the green light will turn on and you will hear the machine operate.  
Immediately seal the area that is being sterilized.  

l.

 

Upon sterilization and scrub cycles completion, the FirstResponder Sterilizer will automatically turn off and the digital display will be 
blank; users can enter the vehicle or room that is being sterilized. If needed, the FirstResponder Sterilizer can be run again with current 
or readjusted settings for further cleaning cycles. 

 

 

TRAINING VIDEO

 

Genlantis provides an online training video for the Ozilla® 
EMS Sterilizer, which is identical in operation and usage as 
the FirstResponder Sterilizer.  The video is available at one of 
these two locations: 
1.

 

On Youtube.com by clinking on, copying and pasting, or 
typing this link: https://youtu.be/DJ8VE8sndoc. 

2.

 

By visiting this Genlantis web page: 
http://www.genlantis.com/ozilla-ems-sterilizer.html and 
clicking on the “VIDEO AND BROCHURE” tab to the right 
of the main product “DESCRIPTION” tab. 

NOTE

: Consult Table 1 on Page 5 for suggestions on 

FirstResponder sterilization times. 

Figure 1:

 Front panel of FirstResponder  Sterilizer 

Содержание FirstResponder E4110FRS

Страница 1: ...g hour mgph 16 000 mg hour mgph De ozonation scrub Fans 4 110 cfm 440 CFM total 4 110 cfm 440 CFM total Warranty 1 year parts and labor 1 year parts and labor Disclaimer The Genlantis FirstResponder Sterilizer is not a replacement or substitute for autoclaves and should not be used to sterilize equipment for use directly in medical or surgical applications of any kind The FirstResponder sterilizin...

Страница 2: ... b Press TIME SELECT button the left most digit will blink c Press SET TIME up or down arrow button repeatedly until desired time is selected d Press TIME SELECT button to move to next digit repeat Step c then press SET TIME to move to the next digit e Repeat Step d one last time to set the last digit the display now should read the ozone time with no blinking digits To set the run time for the sc...

Страница 3: ... s to maximum and run for 3 5 minutes to purge the trapped TO gas c If in a room open windows and let ambient air vent out gradually or use fans to circulate air more effectively TROUBLESHOOTING Problem Solution Unit does not start when OZONE TIME or SCRUB TIME buttons are pressed Check that power cord is attached on the back of the unit Check that the FirstResponder Sterilizer case is closed and ...

Страница 4: ...he terms of this license APPENDIX A WARNING Different materials respond differently to TO molecules and some common first responder gear and triage materials may be damaged by exposure To minimize material damage carefully read and adhere to the following guidelines SE Strong effect Materials are most susceptible to triatomic oxygen TO degradation these materials should be removed from areas to be...

Страница 5: ...emote button to immediately cancel the Ozone cycle and shift the device into the Scrub cycle Allow the FirstResponder device to fully run the Scrub cycle the digital display will go blank before entering space or opening cabinet that is being sterilized Name ________________________________________ Date _________________________________________ Time Sterilization Cycle Started ____________ Expecte...

Отзывы: