OPERATION
1.
Plug power cord into wall outlet, open the door and place objects
to sterilize inside the chamber on either bottom or upper shelf.
Items on upper shelf will be more readily exposed to UV light
radiation, and items on lower shelf will be more exposed to ozone
gas (heavier than air).
NOTE
: because the CoolCLAVE™ BLUE uses UV light and ozone gas,
you are able to safely and effectively sterilize the surfaces of most
types of materials such as plastics, metals, paper products, pipet
tip boxes, and even agar plates. The CoolCLAVE BLUE is not able to
sterilize any liquids or solutions.
2.
Close the CoolCLAVE BLUE door and press the Power Button until it clicks in place; digital display will show “00”.
3.
To select the cycle time, press the Timer Button once (15-min. run time), twice (30 min.) three times (45 min.), and four times
(60 min.).
4.
The Display will show the selected run time, and the UV bulb will automatically turn on and start to count down.
NOTE
: the unit’s door is equipment with a safety switch that will automatically turn the unit off if the door is opened during a
run cycle.
5.
Once the cycle is finished, open the door of the unit while keeping face away so to avoid inhaling ozone gas. Air the chamber
slightly and remove your items for use.
PERFORMANCE
Graph on the right shows performance tests on a CoolCLAVE
Blue Unit. Typical ozone levels expected average 55 ppm.
MAINTENANCE
The CoolCLAVE™ BLUE is a very low maintenance unit and
does not require any scheduled or regular service.
(a)
Keep your CoolCLAVE BLUE unit clean at all times
by wiping the interior and exterior surfaces with a
water dampened towel. To avoid damaging the
unit, do not use any detergents or solvents.
(b)
Unplug the unit from the power outlet when the
CoolCLAVE BLUE is in storage or will not be used for
extended periods of time.
(c)
Avoid damaging the unit by putting heavy objects on top of it.
(d)
Replacing UV-C light Bulbs: The CoolCLAVE BLUE units use a special UV-C light bulb that produce ozone. Please contact
Genlantis to obtain the correct light bulb replacement.
(e)
IMPORTANT: UV bulb waste disposal
: At the end of rated life, when this lamp is removed from service it will be
subjected to the current Toxic Characteristic Leaching Procedure (TCLP) prescribed by the Environmental Protection
Agency. According to TCLP tests these UV lamps are considered hazardous waste under the Universal Waste Rules.
Generators should evaluate all of the disposal options, which may be available in the particular country, state, or city in
which the generator’s facility is located. The generator should check with federal, state and local officials for guidance.
The purchase price paid for the CoolCLAVE™ BLUE grants end users a non-transferable, non-exclusive license to use the unit in a
laboratory or professional
setting, and for sterilizing purposes only
as described in this manual. This use license excludes and without limitation, resale, repackaging, or modification of
the unit in any way and without prior notification of and approval by Genlantis.
Under no circumstances shall the CoolCLAVE BLUE unit be used on food or drink products intended for
human or animal consumption
. At all times, care and
attention should be exercised in handling and using the CoolCLAVE unit by following the instructions in the manual, using common sense lab practices, and
wearing protective clothing and eyewear as needed.
Purchasers may refuse this license by returning the CoolCLAVE BLUE unit unused and its original packaging. By keeping or using the CoolCLAVE unit, you agree
to be bound by the terms of this license. The laws of the State of California shall govern the interpretation and enforcement of the terms of this license.
Genlantis Establishment is registered under Environmental Protection Agency (EPA) ID Number 91845-CA-001