OPERATION 

 

1.

 

Plug power cord into wall outlet, open the door and place objects 
to  sterilize  inside  the  chamber  on  either  bottom  or  upper  shelf.  
Items  on  upper  shelf  will  be  more  readily  exposed  to  UV  light 

radiation, and items on lower shelf will be more exposed to ozone 
gas (heavier than air). 

 

NOTE

: because the CoolCLAVE™ BLUE uses UV light and ozone gas, 

you are able to safely and effectively sterilize the surfaces of most 
types  of  materials  such  as  plastics,  metals,  paper  products,  pipet 
tip boxes, and even agar plates.  The CoolCLAVE BLUE is not able to 

sterilize any liquids or solutions. 

 

2.

 

Close the CoolCLAVE BLUE door and press the Power Button until it clicks in place; digital display will show “00”. 

3.

 

To select the cycle time, press the Timer Button once (15-min. run time), twice (30 min.) three times (45 min.), and four times 

(60 min.). 

4.

 

The Display will show the selected run time, and the UV bulb will automatically turn on and start to count down. 

NOTE

: the unit’s door is equipment with a safety switch that will automatically turn the unit off if the door is opened during a 

run cycle. 

5.

 

Once the cycle is finished, open the door of the unit while keeping face away so to avoid inhaling ozone gas.  Air the chamber 
slightly and remove your items for use.   

 

 
PERFORMANCE 

Graph on the right shows performance tests on a CoolCLAVE 
Blue Unit.  Typical ozone levels expected average 55 ppm. 

 
MAINTENANCE 

The  CoolCLAVE™  BLUE  is  a  very  low  maintenance  unit  and 
does not require any scheduled or regular service. 

 

(a)

 

Keep  your  CoolCLAVE  BLUE  unit  clean  at  all  times 
by wiping the interior and exterior surfaces with a 

water  dampened  towel.    To  avoid  damaging  the 
unit, do not use any detergents or solvents. 

(b)

 

Unplug  the  unit  from  the  power  outlet  when  the 
CoolCLAVE BLUE is in storage or will not be used for 

extended periods of time. 

(c)

 

Avoid damaging the unit by putting heavy objects on top of it. 

(d)

 

Replacing UV-C light Bulbs: The CoolCLAVE BLUE units use a special UV-C light bulb that produce ozone.  Please contact 
Genlantis to obtain the correct light bulb replacement. 

(e)

 

IMPORTANT:  UV  bulb  waste  disposal

:  At  the  end  of  rated  life,  when  this  lamp  is  removed  from  service  it  will  be 

subjected  to  the  current  Toxic  Characteristic  Leaching  Procedure  (TCLP)  prescribed  by  the  Environmental  Protection 
Agency.  According  to  TCLP  tests  these  UV  lamps  are  considered  hazardous  waste  under  the  Universal  Waste  Rules. 
Generators should evaluate all of the disposal options, which may be available in the particular country, state, or city in 

which the generator’s facility is located. The generator should check with federal, state and local officials for guidance. 

 

The purchase price paid for the CoolCLAVE™ BLUE grants end users a non-transferable, non-exclusive license to use the unit in a 

laboratory or professional 

setting, and for sterilizing purposes only

 as described in this manual.  This use license excludes and without limitation, resale, repackaging, or modification of 

the unit in any way and without prior notification of and approval by Genlantis. 
Under no circumstances shall the CoolCLAVE BLUE unit be used on food or drink products intended for 

human or animal consumption

.  At all times, care and 

attention should be exercised in handling and using the CoolCLAVE unit by following the instructions in the manual, using common sense lab practices, and 
wearing protective clothing and eyewear as needed. 

Purchasers may refuse this license by returning the CoolCLAVE BLUE unit unused and its original packaging.  By keeping or using the CoolCLAVE unit, you agree 

to be bound by the terms of this license. The laws of the State of California shall govern the interpretation and enforcement of the terms of this license. 

Genlantis Establishment is registered under Environmental Protection Agency (EPA) ID Number 91845-CA-001

 