SOIL DRYING OVEN 225/425 LITRE 

Page 2 of 3 

SDO 225 & 425 LITRES CAPACITY 

 
 

UNPACKING

 

 

1.1 

Remove all packaging material from between the shelves and inner walls of the oven. 

 
1.2 

Remove all strippable film from the outer surfaces before putting the oven into use. 

 
 

2. 

MAINS SUPPLY 

 

2.1 

Check the voltage and load as shown on the Serial Plate. 

 
2.2 

240 volt units

.  Each unit comes supplied with a mains connection lead already fitted 

with a correctly rated fuse.  The fuse rating and other details for each unit are shown 
on  the  voltage  plate  riveted  to  the  back  of  the  unit.    It  is  important  that,  if  the  fuse 
needs to be replaced, it must ONLY be replaced with one of the correct rating. 

 
2.3 

110  volt  units

  are  supplied  with  a  cable  but  without  a  plug  or  fuse.    These  units 

should be wired in by a suitably qualified electrician to the following:- 

BROWN 

‘L’ Live pin 

Refer to voltage plate 

BLUE 

‘N’ Neutral pin 

for fuse requirement 

GREEN/YELLOW  ‘E’ Earth pin 

 

--- WARNING --- 

DO NOT CONNECT THE OVEN TO A D.C. MAINS SUPPLY 

 
 

3. 

OPERATION 

Note:  As  supplied,  the  controller  is  set  at  107ºC  and  the  over-temperature  selector 
switch set at 120ºC 

 

3.1 

Position the shelves within the work chamber. 

 
3.2 

Switch ‘ON’ the 0-1 mains switch. The oven will now heat up to its present set point 
(107ºC as supplied). 

 
3.3 

To alter the set point, press and hold either the UP or DOWN button for approximately 
5  seconds  -  the  amber  running  LED  will  flash.    You  can  now  set  the  display  to  the 
desired  set  point  temperature  using  the  UP  or  DOWN  buttons.    Once  desired 
temperature  is  entered  and  no  buttons  pressed,  the  display  will  revert  to  the  actual 
temperature  and  the  amber  running  LED  will  stop  flashing  after  approximately  15 
seconds. 

 
3.4 

Over-temperature selector switch – when using the oven at temperatures: 
At or below 110ºC :  have the over-temperature switch in the 120ºC* position 
Above 110ºC 

:  have the over-temperature switch in the 200ºC position 

 

 
*when the selector switch is in the 120ºC position, you can only set the controller to a 
maximum 110ºC 

 
3.5 

If the  chamber  temperature  rises  above  the  overheat  set  temperature,  the  red  LED 
lamp  will  be  illuminated  and  the  heat  control  circuit  will  be  disabled.      Control  will 
switch  back  to  the  main  temperature  controller  once  the  chamber  temperature  falls 
below the overheat set temperature. If the red lamp will not go out or keeps coming 
on there may be a problem with the unit. 

In this case please consult you’re supplier

 