11.3.5
Synchronizing a Master/Slave setup to external time sources
In a Master/Slave setup, all time information is recorded by the Master
mainframe only. Synchronization source selection on Slaves is disabled as
Slave mainframes are dedicated to follow the Master mainframe.
For the Master mainframe, the date and time are controlled by either the PC
(RTC) , PTP master, IRIG or GPS, depending on the synchronization source
selected. The source is selected in the Perception software.
In the
Perception work
area:
Figure 11.11:
Perception work area synchronization source
A
Settings tab
B
General group
C
Sync source column
To set the synchronization source in the Perception software, proceed as
follows:
1
Select the
Settings
sheet.
2
In the
Settings
sheet, go to the
General
group in the task pane and select
Mainframe
.
A list of available mainframes is displayed in the settings area.
3
Select the Master mainframe.
4
Double
-
click on the
Sync source
cell to open it for modification.
5
In the drop
-
down list that appears, select the synchronization source that
should be used.
For information on how to check synchronization, please refer to appendix
"Master/Slave synchronization verification procedure" on page 864.
GEN3t
I3995-3.1 en HBM: public
171
Содержание GEN3t
Страница 1: ...Portable Data Recorder GEN3t English User Manual I3995 3 1 en HBM public...
Страница 4: ...4 I3995 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3t I3995 3 1 en HBM public 51...
Страница 1122: ...1122 I3995 3 1 en HBM public...
Страница 1123: ......