POCKIT™
Central
Nucleic Acid Analyzer
14
Insert the respective sample-loaded Extraction
Cartridge.
Note:
1.
Make sure to remove any residual film from
the Extraction Cartridge surface.
2.
Please refer to user manual of respective
POCKIT™ Central reagent for sample
loading instruction.
3.
Make sure the indicator light of the slot is off
after loading the cartridge.
Enter the
Reagent Lot Number
(stated on the
product label). Press
OK
when completed.
Insert the loaded Transfer Cartridge.
Note:
1.
Make sure to cut excessive film and plastics
outside the rim of Premix.
2.
Make sure the indicator light of the slot is
off after loading the cartridge.
Once the
Reagent Lot No
. is entered, the system
will automatically bring in the channel and target
information.
Note: Please refer to respective user manual of
the POCKIT™ Central reagent to confirm
channel information.
Press
Next
or
Slot B >
to proceed to next slot if you
have more than one sample.
Or press
Confirm
to proceed to
Run Overview
.