![General DTH3007SD Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/general/dth3007sd/dth3007sd_user-manual_2210029006.webp)
(If you press buttons too slowly in setup mode, the screen will revert to the normal display. To
return to the setup sequence, press and hold the SET
▼
button again. To move ahead to the
next field in the sequence or to the next parameter, press the FUNC./HOLD/NEXT button.)
Once you have set the date, the display will prompt you to set the hour, minute and second
of the current time. Again use the SET
▼
or LOGGER
▲
buttons to navigate to the correct
values, and the REC/ENTER button to store the settings.
4. Choose a decimal point or comma
to represent the decimal division between integers
and fractions (for example, American-style 20.88 vs. European-style 20,88). Once you
have set the date and time, the display will show the word “bASIC” above the phrase
“dEC”. Press the SET
▼
button to make “bASIC” (American style) or “Euro” appear in
the upper display, as desired. Press the REC/ENTER button to store the selection as the
default.
5. Enable or disable auto power off
. Once the format of decimal point divisions has been
set and stored, the display will show the word “yES’ over the term “PoFF”. Press the
SET
▼
button until the desired automatic power off management scheme (“yES” for
enable; “no” for disable) is displayed. Press the REC/ENTER button to store the selection.
If enabled, the power off function shuts off the DTH3007SD after a period of inactivity of
ten minutes.
6. Enable or disable the beeper
. Once the auto power off scheme has been chosen and
stored, the display will show the word “yES” over the word “bEEP”. Press the SET
▼
button until the desired setting (“yES” or “no”) is displayed, and then press the
REC/ENTER button to store the selection as the default.
7. Select the thermocouple type
. Once the beep sound has been enabled or disabled,
the lower part of the display will show the term “tYPE”. Press the SET
▼
or LOGGER
▲
button until your thermocouple’s type (K or J) appears on the display. Then press the
REC/ENTER button to store the selection.
8. Select the temperature unit
. Once the thermocouple type has been selected and
stored, the lower part of the display will show the term “t-CF”. Press the SET
▼
or
LOGGER
▲
button until the unit you prefer (“F” for Fahrenheit, “C” for Celsius”) is
displayed. Then press the REC/ENTER button to store the selection as the default.
9. Set the datalogging sampling time
. Once the thermocouple type has been selected
and stored, the display will show a value above the letters “SP-t”. Press the SET
▼
or
LOGGER
▲
button to decrease or increase the value until the desired sampling time
appears above “SP-t”. The options are 0, 1, 2, 5, 10, 30, 60, 120, 300, 600, 1800 and
3600 seconds (0 seconds to 1 hour).
10. Format the SD card
. Once the sampling time has been set and stored, the term “Sd-F”
will appear in the lower half of the display. Press the REC/ENTER button.
Pressing the REC/ENTER button causes the word “no” or “yES” to appear over the term
“Sd-F”. Press the
▼
button to make a selection. Choose “yES” whenever a new SD card is
being used, or when a used card is being repurposed (from use with another SD card
instrument or a camera, for example) and all data on it is to be erased. Choose “no” to
preserve any data on a card previously used with this instrument.