NCIT100 Hawkeye Non-Contact Infrared Thermometer 
Instruction Manual 

Overview 

The Hawkeye Non-Contact Infrared Thermometer is designed for 
fast and accurate body temperature measurements, without 
physical contact. That makes it ideal for mass temperature 
screening and monitoring of individuals at workplaces, public 
events and social gatherings. 

Device Components 

A: LCD backlit screen 

B: Infrared sensor 

C: Measurement Trigger 

D: Battery compartment 

 

Display Screen

 

A: Low battery indicator 

B: °F indicator 

C: °C indicator 

D: Temperature readout 

E: Scan mode indicator 

F: Data hold indicator 

Operating Instructions 

To change the measurement unit

, open the battery compartment 

and move the switch inside the compartment to °F or °C position. 

To install the batteries

, open the battery 

compartment, install 2 

AAA

 batteries according to 

the instruction sticker, and then close the 
compartment door until it snaps shut.  

To take a temperature measurement: 

1.

 

Aim the infrared thermometer (NCIT) at the center of the 
forehead and keep the NCIT perpendicular to the target 
surface. The distance between the NCIT and the forehead 
should be between 0.4 and 1.1 inches.  

2.

 

Press and release the trigger. Dashed lines will appear on 
the screen 

with the word “SCAN” underneath them.

 

3.

 

Do not move the NCIT from this position until the reading 
appears on screen. A reading should appear in under 2 
seconds 

with the word “HOLD” under

neath it. 

4.

 

Depending on the reading, the screen color will change to 
either green, yellow or red and the unit will beep one, two, 
or three times respectively to alert the user to elevated 
temperatures. 

a.

 

98.9°F (37.2°C) and lower, Green & 1 beep 

b.

 

99.0° to 100.3°F (37.3° to 37.9°C), Yellow &  
2 beeps 

c.

 

100.4°F (38°C) and greater, Red & 3 beeps 

5.

 

Repeat this process for each temperature measurement. 
When taking multiple temperatures in a row, if the screen 
seems to be stuck on the last reading wait a few seconds 
and press the trigger again. This should initiate scanning. 

Operation Notes 

1.

 

The NCIT100 provides an oral equivalent temperature based 
on a forehead skin measurement. External sources, such as 
extreme ambient temperatures or physical activity, can 
skew the readings. Keep both the NCIT and the subject in 
the same ambient environment for at least 30 minutes to 
stabilize the device before testing. Multiple readings may be 
necessary to determine an accurate body temperature 
reference. 

2.

 

Make sure the forehead is not covered by anything. Hats, 
hair, sweat, or other obstructions can make the reading less 
accurate. Any obstructions should be removed, and the user 
should wait at least 

10 minutes to take the subject’s 

temperature. 

3.

 

Keep the device away from electro-magnetic fields and 
electrostatic areas. 

4.

 

This device should be used only for screening purposes, and 
not to diagnose any medical conditions.  

Specifications 

Measurement Range: 91.4° to 109.4°F (33° to 43°C) 
Measurement Accuracy: ±0.4°F (±0.2°C) 
Resolution: 0.1° 
Units: °F/°C 
Auto Power Off: 15 Seconds 
Power Supply: 2 

”AAA” Batteries

 

Operating Temperature: 50° to 104°F (10° to 40°C) 

Operating Humidity: ≤

95%RH, non-condensing 

Barometric Pressure: 70kPa to 106kPa 
Storage Temperature: -4° to 131°F (-20° to 55°C) 
Storage Humidity: 

95% RH (w/o battery) non-condensing 

Dimensions: 5.03 x 4 x 1.25in (128 x 102 x 32mm) 
Weight (w/o battery): 0.15lb (70g) 
 

Troubleshooting 

Safety and Maintenance 

1.

 

Clean the infrared sensor lens often with compressed air or 
a moist cotton cloth. Never use a solvent or an abrasive 
cleaner. 

2.

 

The exterior housing (plastic parts only) can be cleaned with 
soap and water on a damp sponge or disinfectant wipe. Do 
not submerge the device or let water get into the interior. 

3.

 

Remove the batteries when cleaning the device. 

4.

 

Remove the batteries before long term storage. 

5.

 

Do not operate the NCIT100 in explosive environments 
containing flammable liquids, gases or dust. 

6.

 

Please dispose of the batteries properly according to your 
local regulations for battery disposal. 

7.

 

Keep out of reach of children and pets. 

8.

 

Do not modify the device in any way. 

 

 

 

Problems 

Causes 

Solutions 

No display 

Dead battery 

Check and replace battery 

 

Low battery 

Replace battery 

Color screen 
dims 

Low battery 

Replace battery 

Reads “OL”

 

The temperature of the 
subject is lower than the 
measurement range 

Make sure the subject is 
within the measurement 
distance 

Reads “OH”

 

The temperature of the 
subject is higher than 
the measurement range 

Make sure the subject is 
within the measurement 
distance 

Reads 

“AL”

 

Ambient temperature is 
too Low 

Make sure ambient 
temperature is within the 
operation temperature range  

Reads 

“AH”

 

Ambient temperature is 
too high 

Make sure ambient 
temperature is within the 
operation temperature range 

Contact Us: 

General Tools & Instruments 
75 Seaview Drive 
Secaucus, NJ 07094 
212-431-6100 
www.generaltools.com 

 

*This device is sold and marketed 

pursuant to FDA’s Enforcement Policy 

for Clinical Electronic Thermometers. 
This device is not FDA 
approved/cleared. 