Chapter 5 Calibration parameters
parameter
Functional description
The
weight of
the
format
unit
A unit used for indicating weight. Initial
value: kg,
Parameters: 0:kg/1: G /2: T /3:lb
The
decimal
point
The number of decimal places used for
weight indication.Initial values: 0
Parameter range: 0/0.0/0.00/0.000
Dividing
the
value
The smallest unit of change in
weight.Initial value: 1.
Parameters: 1/2/5/10/20/50/100/200/500
(d)
range
Maximum
weight
that
can
be
displayed.Initial value: 10000
Parameter range: 1~(100000* degree value)
The
weight of
the
calibrat
ion
Autozero
calibrat
ion
The actual millivolt of the current sensor
is displayed
Input
zero
calibrat
ion
Input the zero millivolt value manually for
zero calibration without weight.
Weight
calibrat
ion
Perform weight calibration parameters
Theoreti
cal
calibrat
ion
Sensor
sensitiv
ity
Average value of the true sensitivity of
the connected sensor, initial value:
2.0000.Scope: 0.0001-5.0000 mV/V
Total
sensor
range
The
sum
of
the
real
range
of
the
sensor;Initial value: 10000
Scope: 0-999999
Enabled
Choose to enable the theoretical value to