Code
Meaning
Corrective Action
E02
No sensor attached
Open the hand held nitrogen purity analyzer and
disconnect and reconnect sensor. Unit should perform an
auto calibration and should read 79.1%. If not, contact
Customer Service for possible sensor replacement.
E02
No valid calibration data available
Make sure unit has reached thermal equilibrium. Press and
hold the Calibration Button for three seconds to manually
force a new calibration.
E02 or ED4
Battery below minimum operating voltage
Replace batteries.
CAL Err St: O
2
Sensor reading not stable
Wait for displayed nitrogen reading to stabilize, when
calibrating the device at 100% oxygen. Wait for unit to
reach thermal equilibrium (Please note that this can take
up to one half hour, if the device is stored in temperatures
outside the specified operating temperature range).
CAL Err lo
Sensor voltage too low
Press and hold the Calibration Button for three seconds
to manually force a new calibration. If unit repeats this
error more than three times, contact Customer Service for
possible sensor replacement.
CAL Err hi
Sensor voltage too high
Press and hold the Calibration Button for three seconds
to manually force a new calibration. If unit repeats this
error more than three times, contact Customer Service for
possible sensor replacement.
CAL Err Bat
Battery voltage too low to recalibrate
Replace batteries.
1-800-345-8207
Calibrating the N2 Analyzer -
G
R 1
alibrate the anal er on initial o er
. herea ter a tec reco
en s calibration on a ee l basis. o serve
as a re in er a one ee ti er is starte ith each ne calibration. t the en o one ee a re in er icon a ears on
the botto o the
. alibration is reco
en e i the ser is ns re hen the last calibration roce re as er or e
or i the eas re ent val e is in estion.
With co
resse air
.1
ne calibration is re ire hen:
he eas re ercentage in .1 is above 0.1
he eas re ercentage in .1 is belo .1
he
re in er icon is blin ing at the botto o the
si
le calibration a be a e ith the sensor o en to static a bient air.
Calibration Errors and Error Codes -
G
R 1
he nitrogen rit anal er has a sel test eat re b ilt into the so t are to etect a lt calibrations o gen sensor ail res
an lo o erating voltage. Re er to the ollo ing table or an e lanation o error co es an ossible actions to ta e.
Содержание NGP-WM300
Страница 2: ...1 800 345 8207...
Страница 7: ...NGP WM lo Diagram 1 800 345 8207...
Страница 8: ...1 800 345 8207 Typical External View Ten allon er al e e ver an...
Страница 21: ...1 800 345 8207 5 o reen...
Страница 32: ...1 800 345 8207 Section 17 Wiring Diagram...
Страница 33: ...1 800 345 8207 6...
Страница 35: ...10 Gallon Receiver Weight 50 lbs 1 800 345 8207 5...