ERRROR MESSAGES & TROUBLESHOOTING GUIDE 

 

ERROR CODE 

 

ISSUE 

 

 

SOLUTION 

Blank Display 

 

Power  

 

 

Check Adapter & Outlet 

ERR-0    

 

 

Overload 

 

 

Capacity Exceeded Or         

 

 

 

 

 

 

 

 

Damaged Weigh Cell 

ERR-2                                Zero Tracking (weight  

Remove All Weight From 

on weigh pan during 

Weigh Pan and Cycle 

power on sequence) 

Power 

 
Stainless Steel Weigh 

Reassemble with Weigh 

Pan Not Assembled at 

Pan, Sub Pan and 

                                            Calibration                         Calibrate 
 
ERR-5 

 

 

Internal Signal 

  

Weigh Pan Alignment 

Unstable 

 

 

Ensure Weigh Pan & Draft  

 

 

 

 

 

Ring Placement are Correct 

 
 

 

 

 

Unstable Power – Check    

 

 

 

 

 

Power Supply / Transformer  

   

 

 

 

ERR-C 

 

Calibration Error 

 

Repeat Calibration Process 

 

Incorrect Weight 

 

Ensure Correct Calibration Weight

 

 

 

 

 

Is Used – See Page 3 

 
ERR-L 

 

 

Calibration Error 

  

Repeat Calibration Process 

 

 

No Weight Placed  

 

 

On Weigh Pan 

 
ERR-E 

 

 

Memory Error 

  

Cycle Power (On/Off Switch) 

 

:      

 

 

 Low Battery   

 

Replace Batteries 

 Indicator 

 

 

6 AA. Replace all  

 

 

 

 

Batteries at Same Time 

 

                  

Page 6 

 

Содержание GP-Series

Страница 1: ...GP Series Portable Balances Operation Manual ...

Страница 2: ...BATTERY OPERATION 2 LEVELING 3 CALIBRATION 3 WEIGH MODE SETTINGS 3 BACKLIGHT SETTINGS 4 AUTO OFF SETTINGS 4 TARE FUNCTION 5 CARE MAINTENANCE 5 ERROR MESSAGES TROUBLESHOOTING GUIDE 6 SERVICE TECHNICAL SUPPORT 7 DISPOSAL 7 CE COMPLIANCE MARKING 8 SPECIFICATIONS 9 FEATURES 9 INSTRUMENT DIAGRAM 10 i ...

Страница 3: ... or dropping samples onto the weigh pan Weight overload and shock may damage the instrument and void the warranty When utilizing the battery operation mode do not mix batteries Replace all six batteries at the same time If the instrument will not be used for an extended period of time remove all batteries from the instrument Failure to do so may cause leakage and damage to the instrument GETTING S...

Страница 4: ...and ambient temperate conditions require calibration at the point of use Calibrate regularly at least once a week to ensure accurate weighing results AC POWER CONNECTION BATTERY OPERATION OPTION When connecting power use the original AC adapter supplied with the balance Using an unapproved power supply may damage the instrument and void the warranty Connect the plug to the instrument Ensure the po...

Страница 5: ...ght for your balance see chart below will flash on the display 5 Place calibration weight on the weigh pan 6 When calibration is complete the balance will display the selected calibration weight value beep and return to the weigh mode Calibrate on a regular basis we recommend once a week or more Calibration Weight Values B 1852 120 100g ASTM Class 3 B 1852 500 500g ASTM Class 4 B 1852 1500 1000g A...

Страница 6: ...off Press ON OFF again to turn on the power and the scale return to the weighing mode AUTO SHUTOFF SETTINGS All GP Series balances come with an AUTO OFF feature typically used in the battery operation mode The AUTO OFF feature is deactivated at the factory However should you wish to activate the Auto Off feature Press ON OFF Key to turn the power off press again to turn power on When instrument di...

Страница 7: ... the reading is 0 before placing a sample on the weigh pan If not press TARE KEY to display 0 CARE MAINTENANCE Unplug the instrument from the power source and remove all batteries from the battery compartment Wear appropriate safety protection Clean inside hood when appropriate A damp cloth is recommend for basic cleaning Do not use solvents on the instrument housing sub pan keypad power cord and ...

Страница 8: ...sembled at Pan Sub Pan and Calibration Calibrate ERR 5 Internal Signal Weigh Pan Alignment Unstable Ensure Weigh Pan Draft Ring Placement are Correct Unstable Power Check Power Supply Transformer ERR C Calibration Error Repeat Calibration Process Incorrect Weight Ensure Correct Calibration Weight Is Used See Page 3 ERR L Calibration Error Repeat Calibration Process No Weight Placed On Weigh Pan ER...

Страница 9: ...hat has been exposed to biological or hazardous material contamination for replacement credit repair or disposal DISPOSAL Please consider the environment when disposing of your instrument and the packing material Please recycle all environmentally friendly waste Please contact your local government agency facilities manager or a commercial disposal operator on the proper disposal of the instrument...

Страница 10: ...Limits Applicable European Standards EN 60950 Safety Requirements for electrical equipment for measurement control and laboratory use Part1 General requirements Note Modification of this instrument in any manner is the sole responsibility of the owner operator In addition to voiding the warranty the owner operator is responsible to check and if necessary correct any modifications required in accor...

Страница 11: ...tabilization Time 3 Seconds Operating Temperature 0C to 40C 32 104F Power Requirements Input 100 240VAC 50 60Hz 0 4A Output 9VDC 300mA Battery Operation 6 AA Batteries not included Unit Dimensions B 1852 120 225 x 177 x 124mm 8 9 x 6 7 x 6 2in All Other Models 225 x 177 x 70mm 8 9 x 6 7 x 2 8in Weigh Bowl 200 x 200 x 55mm 7 9 x 7 9 x 2 2in Net Weight 2 kg 4 4lb FEATURES Large Backlit Display Auto ...

Страница 12: ...INSTRUMENT DIAGRAM LEFT SIDE AC ADAPTER INSERT WEIGHT DISPLAY ON OFF UNIT CAL TARE BOTTOM BATTERY COMPARTMENT Batteries are not included DISPLAY LOW BATTERY INDICATOR Page 10 ...

Страница 13: ......

Отзывы: