button, while turning on the unit with the button
.
Scroll through the menu options using the (
) button until the
following message is displayed, then select by pressing the
button.
X-level
Rezero?
When prompted with the following message, place the Garnet
firmly and squarely on the track, making a note of its exact
location.
Measure
A--
==
--B
Press the
button to begin the calibration process. During the
measuring process a ‘thermometer’ bar graphs indicates the
progress. It is essential that the Garnet is not touched during this
measuring process
Processing…
When prompted with the following message, turn the Garnet
around so the Gauge paddle is now on the opposite rail but at
EXACTLY the same location on the track
Rotate 180?
B--
==
--A
Ensure that the Garnet is firmly and squarely positioned and press
the
button to continue the calibration process. During the
measuring process a ‘thermometer’ bar graphs indicates the
progress. It is essential the Garnet is not touched during this
measuring process.
Processing…
Once the calibration process is complete, the following message
will be displayed. If Cross-level Rezeroing was initiated manually
by selecting from within the ‘user menu’ then the number displayed
represents the ABSOLUTE offset for Cross-level. If however the
rezeroing was initialised automatically (due to a change in
temperature) then the number displayed represents the ‘adjustment’
made to the previously stored value, thus giving an indication of the
previous offset error.
Calibrated:
+0.4
m
m
At the end of the Cross-level Rezeroing process the final value is
checked to ensure it is within certain criteria. If these criteria are
not met then the calibration is aborted and the following message is
displayed. In this instance, the user should attempt to re-zero again
before using the Garnet.
Failed!
GAD04DOC820Ver9 - en.docx