8555-9201-000
/
0109
71
Mechanical Separation /
GEA Westfalia Separator
4.26 Alarm #39 - PT1 function (option)
PT1 function
In clean oil discharge
Separation “Stop”
Alarm#39 ENTER
Text display
Possible causes
Action
No pressure drop at PT1 during the
ejection.
•
Check switching function.
Pt1 is soiled, catches or is defective.
•
Clean and check PT1.
PT1 is set too low.
•
Check switching point PT1.
No ejection, ejection monitoring has
not responded.
•
Check bowl hydraulics and ejection
monitoring.
System pressure too high in clean oil
discharge (product discharge).
•
Check delivery head at clean oil
discharge (product discharge).
•
Deselect function if required.
Help text – information on the control system
E N T E R
Step 1
Acknowledge fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
PT1 FUNCTION
Please check:
Switching function PT1
PT1 soiled
->
Page 2
Switching point PT1
Bowl hydraulics
Ejection monitoring
Delivery head at
->
Page 3
clean oil discharge
back with ENTER
F 1