Chapter 7 - Diagnostics/Troubleshooting
7-49
7-7-3-3
DRFG Overall Test
This diagnostic test performs the self-diagnosis of each electric component and interface in the DRFG
board.
below outlines Voluson™ P8/Voluson™P6 Diagnostics Tool described in the sub-sections.
7-7-3-3-1
Voltage DRFG
It verifies Each Power Voltage Level on DRFG board, and displays the current voltage levels and the
test result as PASSED/FAILED.
7-7-3-3-2
Temperature DRFG
It verifies each GPU, CPU, DPS temperature measure on DRFG and displays the current temperature
status and the test result as PASSED/FAILED.
7-7-3-3-3
IIC DRFG Interface
It verifies I2C path named I2C_Local by using EEPROM on DRFG board and displays the test result as
PASSED/FAILED.
7-7-3-3-4
HW ID Config
It verifies HW ID configuration on DRFG board and displays the current HW ID and the test result as
PASSED/FAILED.
7-7-3-3-5
FPGA DRFG Interface
It verifies FPGA(DRFG) configuration on DRFG board by reading/writing test data and displays the test
result as PASSED/FAILED.
Table 7-3
Automatic diagnostic test under DRFG Test
Sub-section
Description
Page Number