Section 8-57 - Replacement of ECG Kit assy
Section 8-57
Replacement of ECG Kit assy
8-57-1
Manpower
One person, 10 minutes.
8-57-2
Tools
Philips screwdriver
8-57-3
Preparations
Power Off/Shutdown the system as described in
.
8-57-4
Removal Procedure
1.) Remove 5 probe holders.
2.) Unscrew 9 numbers of screws below the keyboard base.
Figure 8-254 Removing 5 Probe holders and unscrewing 9 screws below the keyboard base