GE
RAFT
V
OLUSON
™ P8 / V
OLUSON
™ P6
DIRECTION 5723243, R
EVISION
6
DRAFT (M
AY
23, 2018)
B
ASIC
S
ERVICE
M
ANUAL
7-32
Section 7-7 - Hardware Tests
7-7-3-3
DRFM Overall Test
This diagnostic test performs the self-diagnosis of each electric component and interface in the DRFM
board.
below outlines Voluson™ P8 / Voluson™ P6 Diagnostics Tool described in the sub-sections.
7-7-3-3-1
Voltage DRFM
It verifies Each Power Voltage Level on DRFM board, and displays the current voltage levels and the
test result as PASSED/FAILED.
7-7-3-3-2
Temperature DRFM
It verifies each GPU, CPU, DPS temperature measure on DRFM and displays the current temperature
status and the test result as PASSED/FAILED.
7-7-3-3-3
IIC DRFM Interface
It verifies I2C path named I2C_Local by using EEPROM on DRFM board and displays the test result as
PASSED/FAILED.
7-7-3-3-4
HW ID Config
It verifies HW ID configuration on DRFM board and displays the current HW ID and the test result as
PASSED/FAILED.
7-7-3-3-5
FPGA DRFM Interface
It verifies FPGA(DRFM) configuration on DRFM board by reading/writing test data and displays the test
result as PASSED/FAILED.
Table 7-2
Automatic diagnostic test under DRFM Test
Sub-section
Description
Page Number