7.4.2 Export Log's and System Data
There are two possibilities to export system data (and log´s):
1.
by pressing the Alt + D key to save a snapshot and “Full Backup” of the System state; see
2.
via the
Export System Data
button in the System Setup - Administration - Service page; see
7.4.2.1 Export System Data (by pressing the ALT + D key)
Alt + D uses “Full Backup” to gather data from the system. In addition it creates one screen shot (Alt-D.bmp)
of the point in time when Alt + D was pressed. The main use is when R&D or OLC need detailed information
about the system (e.g., when experiencing strange behaviour or when the problem should be investigated by
R&D). It is not intended to replace or enhance the existing Full Backup functionality.
The Full Backup created by Alt + D is protected by a password that can be customized. Whenever
transmitting system state to R&D, do not forget to inform them about any password change.
Data can be stored on the hard disk (D:\export\fullbackup\fb1), or you can export them to DVD/CD, etc.
Including the D:\export folder, which contains dump files (for details see
Section 7.4.2.2.1 on page 7-11
Autotester script files, SMART logs, sniffer logs and screen shots (Alt + P).
Note
Successive Alt + D keystrokes overwrite existing snapshots at destination (Internal) HDD.
1.
Press the Alt + D key on the keyboard simultaneously.
Figure 7-9 "Alt-D" backup
2.
Select whether you want to backup data with our without hardware diagnosis tests.
Please detach all probes if you want to execute hardware diagnosis, then click
Yes
.
Figure 7-10 System Configuration Save
3.
Select the destination of the “System state” backup.
4.
Select the
Next
button to start the backup process.
After saving the data, the Voluson E-Series reboots and the application starts again.
Diagnostics/Troubleshooting
7-10
Voluson E-Series Service Manual
5539550APB Revision 6
Содержание Voluson E Series
Страница 11: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 13: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 15: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 17: ...Introduction Voluson E Series Service Manual 5539550APB Revision 6 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 120: ...This page was intentionally left blank Setup Instructions 3 76 Voluson E Series Service Manual 5539550APB Revision 6 ...
Страница 355: ......
Страница 356: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...