Section 8-6 - Lower Section Components Replacement Procedures
P R E L I M I N A R Y
8-6-14
DIB (Distribution Interface Board) Replacement Procedure
8-6-14-1
Tools
Use the appropriate flat and Phillips-type screwdrivers and a wire cutter as indicated in the
DIB replacement procedure.
8-6-14-2
Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in
.
8-6-14-3
DIB Removal Procedure
1) Remove the system left and right side covers, as described in the
8-2-3-3 "Left Side Cover Removal
8-2-2-3 "Right Side Cover Removal Procedure" on page 8-5
.
2)
Remove the system front cover, as described in the
8-2-6-3 "Front Cover Removal Procedure" on
3.) Disconnect any peripherals currently connected to the system.
4.) Remove the Peripherals AC Outlet Connectors Panel, as described in
Outlet Connectors Panel Removal Procedure" on page 8-167
.
5) Remove the sub-woofer, as described in the
8-6-13-3 "Sub-woofer Removal Procedure" on page
6) Unplug each of the six internal cables from the DIB Box, as shown in
NOTICE
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to
8-6-2-4 "Battery Removal Procedure" on page 8-116
Figure 8-218 Cables Connected to DIB Box
DVI DVR
(J20)
USB Printer
(J15A)
USB
(J14A)
Display PWR
(J11)
AC Box
(J7)
KB Data
(J1)
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...