2-2
Section 2-2 - Console Requirements
P R E L I M I N A R Y
Section 2-2
Console Requirements
2-2-1
Unit Environmental Requirements
NOTE:
The Vivid™ S5 N and Vivid™ S6 N system may be operated at an altitude of up to 3000 meters
(9842 ft)
NOTE:
After a long period of storage, or after transportation of the system with the monitor in the folded-down
position, it is highly recommended to place the monitor in the upright position - and to leave it in this
position for a period of longer than 1 hour before use. This will enable it to properly adjust to the
environmental conditions.
2-2-2
Cooling Requirements
The cooling requirement for the Vivid S5 N or Vivid S6 N ultrasound unit environment is 3500 BTU/hr.
This figure does not include the cooling required for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/hr demand on the environmental
cooling.
2-2-3
Lighting Requirements
For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being performed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that monitor
lighting can be adjusted automatically on the Vivid S6 N if this option is used.
Table 2-2
Environmental Requirements
Requirement
Temperature
Relative Humidity
(non-condensing)
Air Pressure
Operational
10 —
40
o
C (50 — 104
o
F)
30 — 85%
700 — 1060 hPa
Storage
-20 — 60
o
C (-4 — 140
o
F)
30 — 70%
700 — 1060 hPa
Transport
-20 — 60
o
C (-4 — 140
o
F)
30 — 70%
700 — 1060 hPa
CAUTION
If the system has been in storage or has been transported, please see the acclimation
requirements
before
powering
ON
and/or using the system.
3-2-2 "Installation Warnings" on
.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...