Section 4-2 - General Procedures
P R E L I M I N A R Y
4-2-5
Using Removable Media
4-2-5-1
Introduction
Vivid S5 N and Vivid S6 N ultrasound scanners are equipped with a Magneto Optical Disk Drive
(MO Drive) and a DVD/CD-RW Drive.
The MOD Drive can be connected to either of the USB ports (
upper
or
lower
) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor being connected to
the Vivid S5 N or Vivid S6 N USB port, and the other side to the rear of the MOD drive.
When connecting the MOD drive directly to the system, it is necessary to use an additional power
source.
Dedicated peripheral power outlets are available at the rear of the system, providing
power to the system’s peripherals via an internal insulation transformer.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed designated to be used with the specific MOD
drive supplied by GE HEALTHCARE.
4-2-5-2
Install MO Disk in MO Drive
1.) Preparations before installing an MO disk in the MO Drive:
a.) The unit must be turned ON and “up and running” before using the MO Drive
b.) Verify the MO disk for loose hardware or damaged labels which could jam inside the MO Drive.
c.) Ensure that the slide switch in one corner of the disk is set so that the disk is write enabled
(disk hole closed).
2.) Insert the disk into the MO Disk Drive with the label for the side you intend to use, facing up.
NOTE:
A new physical storage message is displayed.
4-2-5-3
Eject and Remove MO Disk from MO Drive
In some cases, the MO Disk is ejected automatically. If not automatically ejected, use the manual
ejection method as described in the following steps:
1.) Press the
Eject
button on the keyboard
OR
Press <
Alt> + E
on the alphanumeric keyboard to display the Eject Device menu.
WARNING
WARNING
Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive
and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
NOTICE
Never move the unit with a MO disk in the Magneto Optical Disk Drive because the drive actuator
will not be locked and the Magneto Optical Disk Drive could break.
NOTICE
Do not eject the MO Disk using the button on front of the MO drive.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...