Chapter 3 - System Setup
3-141
P R E L I M I N A R Y
3-7-8
eVue Dataflow Configuration
NOTE:
MPEGVue Player is intended for secondary review and patient consultation.
3-7-8-1
Introduction
It is possible to configure the Vivid S5 N or Vivid S6 N ultrasound scanner to allow the user to send
images or loops to a remote PC, in almost real time. When pressing Store, the loop will save not only
to the local archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”)
on the network (in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital
in real-time
is required.
3-7-8-2
Preparations
Before configuring the Vivid S5 N or Vivid S6 N for eVue dataflow, it is necessary to decide upon a
specific PC in the hospital as being the e-Vue PC station. The service technician then configures the
Vivid S5 N or Vivid S6 N to send images to that specific e-Vue PC, as described below.
3-7-8-3
Configuring the Vivid S5 N or Vivid S6 N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid S5 N or Vivid S6 N is connected to the Network via cable or wireless.
3.) On the Vivid S5 N or Vivid S6 N scanner, press the
Config
button.
The System Configuration window opens.
4.) Click
Connectivity
.
The Connectivity dialog box opens.
5.) Select the
Dataflow
Tab, as shown in
.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...