Chapter 3 - System Setup
3-99
3-10-3-3
Set the Scanner’s Network Information
In the IP settings area of the screen, enter the following:
1.) IP Address for scanner. (Default IP Address from factory: 10.0.0.3).
2.) Subnet Mask for scanner. (Default Subnet Masks from factory: 255.255.255).
3.) IP address for Default Gateway. (Default Gateways from factory IP address: 0.0.0.0).
4.) In addition, the scanner’s AE Title must be entered in the DICOM server’s setup.
.
3-10-3-4
Set the Remote Archive’s Network Information
In the Remote Archive Setup area of the screen (see example in
), enter the following:
1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Archive Name from factory: ECHOPAC7-000001).
.
Figure 3-101 TCP/IP Set-up for Vivid q N
Figure 3-102 TCP/IP Set-up for Vivid q N
IP Address
Subnet Mask
Default Gateway
AETitle
IP Address for the remote archive
Computer Name for the remote
archive
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...