GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
7 - 20
Section 7-9 - Troubleshooting
7-9-9
System Temperature Too High
When the temperature inside the VIVID E9 increase, the fan speed will increase to cool down the
system. If the air filters become too dusty, a higher fan speed is required to keep the air stream at the
needed level. When the fan speed increase, the fan noise will also increase.
•
If the cooling air stream is insufficient to stabilize the interior temperature within the operating
margin, the system will stop operating.
-
Solution
: Clean or replace the filters.
•
If a fan is worn out, it may be noisy, or stop working.
-
Solution
: Replace the fan(s).
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......