Replacement Procedures
8-108
Vivid E80/E90/E95
–
Service Manual
GC091052
Rev. 3
3. Remove the two fastening screws on the Patient I/O
module’s mounting bracket.
4. Pull the Patient I/O module sideways, out of the frame.
Patient I/O installation procedure
Follow these steps to install the Patient I/O:
1. Align the Patient I/O with the opening and push it into its
final position.
2. Connect the Power and USB cables (inside the BEP).
3. Install two fastening screws on the mounting bracket.
4. Install the BEP Cover.
5. Install the left side cover.
Verification
Perform the following steps to verify that the product is
functioning as intended after this Patient I/O replacement:
1. Verify that all screws that you removed earlier have been
installed.
2. If finished, connect cables and probes you removed earlier.
3. Power up the Ultrasound system to verify that it operates as
intended.
CAUTION
Be careful not to pinch any of the cables.
Содержание Vivid E80
Страница 4: ...i 2 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 Important precautions Translation policy ...
Страница 5: ...Vivid E80 E90 E95 Service Manual i 3 GC091052 Rev 3 ...
Страница 6: ...i 4 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 7: ...Vivid E80 E90 E95 Service Manual i 5 GC091052 Rev 3 ...
Страница 8: ...i 6 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 9: ...Vivid E80 E90 E95 Service Manual i 7 GC091052 Rev 3 ...
Страница 10: ...i 8 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 11: ...Vivid E80 E90 E95 Service Manual i 9 GC091052 Rev 3 ...
Страница 12: ...i 10 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 13: ...Vivid E80 E90 E95 Service Manual i 11 GC091052 Rev 3 ...
Страница 14: ...i 12 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 15: ...Vivid E80 E90 E95 Service Manual i 13 GC091052 Rev 3 ...
Страница 16: ...i 14 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 22: ...i 20 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Страница 30: ...i 28 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Страница 54: ...Introduction 1 24 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Страница 74: ...Site Preparations 2 20 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Страница 318: ...Replacement Procedures 8 96 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 Back End Processor BEP parts replacement ...
Страница 336: ...Replacement Procedures 8 114 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Страница 371: ...Vivid E80 E90 E95 Service Manual 10 1 GC091052 Rev 3 Chapter 10 Care and Maintenance ...
Страница 376: ...Care and Maintenance 10 6 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Страница 381: ......