Chapter 8 - Replacement Procedures
8 - 103
Section 8-19
Brake Pedal and Direction Lock Assembly Replacement Procedures
8-19-1
Manpower
One person, 1 hour
8-19-2
Tools
•
Phillips screwdriver size 2
•
Hexagon key 5 mm
•
Torx screwdriver T-20
•
Wrench 11 and 13 mm
8-19-3
Preparations
"Front Bumper Removal Procedure" on page 8-100
2.) Remove footrests on both pedals by unscrewing 4 Torx screws on each.
8-19-4
Brake Pedal and Direction Lock Removal Procedure
1.) Unscrew 2 nuts on the two ball joints on the brake rods, located over the middle of brake assembly.
Wrench 11 and 13.
2.) Unscrew the 4 hexagon screws. Two of them located under the Brake Assembly.
NOTE:
The nuts may slide out of position.
3.) Remove the Brake Pedal and Direction Lock Assembly.
Figure 8-100 Brake Assembly
Hexagon screws
Ball joint nuts