Section 5-10 - System Power Distribution
P R E L I M I N A R Y
Section 5-10
System Power Distribution
5-10-1
Introduction
The Venue™ system power distribution consists of the following modules:
•
An Isolated AC/DC Power supply to all System modules (18v/400W) called the PSU Unit.
•
Optional Battery Packs that supplies Power to the system when AC power line is not present or in
the event of AC power failure (12-16.8V).
•
Input filter unit, IPP, before PSU, for filtering AC line noises.
5-10-2
AC Distribution
The PSU module converts the AC power line (100-240V AC) to 18V DC @ 400 W. this module is the
main power supply for the Venue™ ultrasound scanner:
5-10-3
DC Power Distribution
The main DC Power supplied from the PSU (AC/DC) is distributed by the MPB to the following system
modules:
•
Back End Module.
•
Front End Power Supply (T-FEPS).
•
Cockpit (Monitor)
•
Battery charging
•
Peripherals