![GE Venue 50 Скачать руководство пользователя страница 58](http://html2.mh-extra.com/html/ge/venue-50/venue-50_basic-service-manual_4174021058.webp)
Site Preparations
2-6
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Electrical requirements
General requirements
NOTE:
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet
.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet
.
NOTE:
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
Electrical requirements for the Ultrasound system
Inrush Current
Inrush current is not a factor to consider due to the inrush
current limiting properties of the power supplies.
Table 2-3: Electrical Specifications for
Venue 50 system
Voltage
Power
Current
Frequency
100-240 VAC
180 VA max.
1.6A (max)
50/60HZ
Table 2-4: Electrical Specifications for
Docking Cart
Voltage
Power
Current
Frequency
100-240 VAC
380 VA max.
2.7A (max)
50/60HZ
Table 2-5: Inrush Current
Voltage
Inrush Current
100V
75A (max)
240V
120A (max)
Содержание Venue 50
Страница 4: ...i 2 Venue 50 Basic Service Manual 5447566 100 English Rev 7 Important Precautions Translation policy...
Страница 5: ...Venue 50 Basic Service Manual i 3 5447566 100 English Rev 7...
Страница 6: ...i 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 7: ...Venue 50 Basic Service Manual i 5 5447566 100 English Rev 7...
Страница 8: ...i 6 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 9: ...Venue 50 Basic Service Manual i 7 5447566 100 English Rev 7...
Страница 10: ...i 8 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 11: ...Venue 50 Basic Service Manual i 9 5447566 100 English Rev 7...
Страница 12: ...i 10 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 22: ...i 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 52: ...Introduction 1 30 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 72: ...Site Preparations 2 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 156: ...General Procedures and Functional Checks 4 28 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 160: ...Components and Functions Theory 5 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 177: ...Troubleshooting Venue 50 Basic Service Manual 7 11 5447566 100 English Rev 7 Noise in B Mode Figure 7 9 Noise in B Mode...
Страница 204: ...Troubleshooting 7 38 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 228: ...Replacement Procedures 8 24 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 284: ...Care and Maintenance 10 36 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 288: ...Index 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7...
Страница 289: ......