Performing a scanning 4.4
Typhoon instrument states
The Typhoon instrument can be in one of five states. The
Typhoon
Scanner Control
window
Instrument State
area (A) displays the current
instrument state (
Warming Up
,
Ready
,
Sleep
,
Initialization
or
Scanning
) and an
Initialize Scanne
r button.
1
Warming Up
The
Warming Up
message is shown for 5 minutes after you turn on the system or
click the
Initialize Scanner
button.
Note:
Complete warm-up takes 30 minutes when the instrument is turned on.
2
Ready
The instrument is ready to scan.
Note:
If you just turned on the instrument, the Ready message appears after
approximately 5 minutes. However, for best performance, the instrument
should be warm up for the full 30 minutes before scanning.
3
Sleep
The lasers have been shut off and the
Sleep
state message appears. This happens
when the instrument has not been used for the preset time (default setting is 4
hours, change time by using the Typhoon Direct Instrument Access software).
4
Initialization
After you click the
Scan
button, the
Initialization
state message appears. It
remains displayed until the instrument begins scanning.
Initialization
state takes approximately 1 minute if the prior instrument state was
Ready
, and approximately 5 minutes if the prior instrument state was
Sleep
or
Warming Up
.
A
Содержание Typhoon 9200
Страница 2: ...GE Healthcare Typhoon Operating Instructions Original instructions...
Страница 3: ......
Страница 13: ...1 Introduction 1 4 Control software 12 Typhoon Operating Instructions 28 9623 56 AA...
Страница 27: ...3 Installation 3 5 Spare parts and accessories 26 Typhoon Operating Instructions 28 9623 56 AA...
Страница 75: ...Appendix A Accessories A 4 Beamsplitters 74 Typhoon Operating Instructions 28 9623 56 AA...
Страница 76: ......