![GE TX-6310-01-1 b Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/ge/tx-6310-01-1-b/tx-6310-01-1-b_manual_4107871002.webp)
•
Do not install near ceiling fans, doors, windows, or areas
directly exposed to the weather.
•
Do not install in dead air spaces, such as peaks of vaulted
ceilings or gabled roofs, where CO may not reach the sensor
in time to provide early warning.
•
Do not install near deep-cell large batteries. Large batteries
have emissions that can cause the alarm to perform at less
than optimum performance.
•
Do not obstruct the vents located on the alarm. Do not place
the alarm where drapes, furniture, or other objects block the
flow of air to the vents.
Battery installation and replacement
Note:
Place the control panel into sensor test mode prior to
replacing the batteries. If the control panel is not in sensor test
mode during battery replacement, an alarm/tamper condition
may be indicated.
To install or replace the batteries in this alarm:
1. Slide the alarm body off of the mounting plate.
2. If replacing batteries, remove the old batteries and properly
dispose of them as recommended by the battery
manufacturer.
3. Install the new batteries. Note the polarity illustration in the
battery compartment (see Figure 3 below).
Figure 3: Batteries
W
A
R
N
IN
G
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Battery compartment
_
+
_
4. Slide the alarm body back onto the mounting plate.
Note:
The mounting plate will not close if all three batteries
are not installed.
Figure 4: Sliding the alarm on the mounting plate
Alarm mounting guide
Alarm mounting guide
5. Perform a sensor/RF test with the control panel. See “RF
When replacing the batteries, use one of the following approved
brands:
•
Duracell MN1500 or MX1500
• Energizer
E91
Use of a different battery may have a detrimental effect on the
alarm operation.
Note:
Constant exposures to high or low humidity may reduce
battery life.
After installing or changing the batteries, reinstall your alarm. Test
your alarm by using the Test/Hush button and check that the
green Power LED is on.
Mounting the alarm
Note:
Verify RF performance prior to permanently mounting the
alarm. See “RF communication test” on page 3.
The CO alarm can be wall mounted or ceiling mounted.
To mount the alarm:
1. Slide the alarm body off of the mounting plate. Place the
mounting plate in the desired location, and mark the location
of the two mounting holes. Orient the mounting plate
vertically or horizontally as shown in the following figures.
Note:
The alarm can also be directly mounted to a single
gang box.
Figure 5: Vertical mounting
Mounting hole
Mounting latch
Alarm
procedure
label
Warning
XXXXxx XXX x xxx
XX XXX XXXXxxx
XXXX XXX XXXxx
XXXXXXXX XX xx
XXXXxx XXX x xx
XX XXX XXXX xx
XXXX XXX XXXxx
XXXXXXXX XX xx
XXXXxx XXX xxx
XX XXX XXXXxx
XXXX XXX XXXx
XXXXXXXX XX
XXXXxx XXX xxx
XX XXX XXXX xx
XXXX XXX X X
XXXXXXXX XXxx
xxxxxxxxxxx
XXXXxxx
xxxxxxxxxxx
Mounting hole
2 / 8
P/N 466-2393 • REV 00.15 • ISS 07DEC10
Draft