Cleaning, Maintenance, Disposal
28
TONOPORT VI
2001589-312 Revision B
Technical Inspections of the Measuring Sys-
tem
The non-invasive pressure measurement system of
TONOPORT VI should be inspected every two years.
The checks can be carried out by GE Healthcare within
the framework of a service agreement; please contact GE
Healthcare Service for details.
The nature and scope of these checks are explained in the
corresponding sections of the Field Service Manual.
On request, GE Healthcare
will provide a Field Service
Manual.
Disposal of the Product
Calibration Mode
(e.g., to check the pneumatic system for leaks)
Connect a rubber bulb between pressure tubing and
cuff, using a T-adapter.
Roll up cuff tight.
Turn device off and on again after a few seconds, then
wait for time to be displayed.
Push
four times: the display indicates "H 4".
Push
: the display indicates an internal value
that must be between 25 and 100. If the displayed
value is outside this range, TONOPORT VI must be
returned for repair.
Push
again: the display indicates "0" (the
display now indicates the pressure in mmHg).
Generate a test pressure of 200 mmHg and measure
the pressure decrease after waiting at least 30
seconds. (Pressure decreases between 3 and 5 mmHg
are typical; a pressure decrease > 6 mmHg indicates a
leak and the system needs to be repaired.)
Push
to exit the calibration mode.
Viewing the Firmware Version
Turn on the device and wait for the time to be
displayed.
Push
five times: the display indicates "H 5".
Push
: the firmware version is indicated, e.g.,
– "30" =
firmware version 3.0
Push
to end the display.
Caution
These checks shall be carried out by GE Healthcare
or authorized companies.
The product described in this operator
manual must not be disposed as unsorted
municipal waste and must be collected
separately. Please contact an authorized
representative of the manufacturer for
information concerning the
decommissioning of your equipment.
INFO
START
STOP
START
STOP
START
STOP
INFO
START
STOP
START
STOP