GE Telaire Vaporstat 9002 Скачать руководство пользователя страница 5

Telaire Vaporstat™ Model 9002

Page 5

Zero Point Calibration

All tubing should be connected between the gas bottle and the sensor 

inlet flow port, as shown in Figure 2 below. Gas should be flowing to the 
sensor at a rate of 80-100 cc/minute for a minimum of 5 minutes prior to 
initiating calibration.

Figure 2: Zero Point Calibration - Setup

1.

Attach the short hose to the bottom port on the flowmeter. 

2.

Attach the long hose to the top port on the flowmeter.

3.

To insure the flowmeter is kept in the vertical position, secure the 

flowmeter to the side of gas bottle using the supplied tie wrap.

4.

Remove the protective cap from the nitrogen bottle and attach the 
gas regulator.

5.

Attach the open end of the bottom hose (located on the flowmeter) to 
the gas regulator. Slide the hose far enough on the gas port to ensure 

a secure, airtight connection.

6.

Insert the male Luer fitting (located on the longer hose) into the 
calibration port, located on the bottom of the Model 9002.

7.

Verify all the components are installed correctly and initiate the 
calibration process by turning the knob on the regulator. Turn the 

knob until the indicator reaches 7 psi.

8.

Turn the flowmeter knob until the floater reaches 80 -100 cc/ min.

9.

Allow the gas to flow for at least 5 minutes before proceeding.

10.

Press and hold the 

Clear + Enter

 buttons on the keypad for 5 

seconds. The word 

CAL

 will appear in the top line of the LCD 

display.

11.

Use the 

Arrow

 buttons to toggle to the 

ZERO

 calibration mode.

12.

When 

ZERO

 is displayed, press the enter button to initiate the 

calibration process.

13.

Once the 

Enter

 button is pressed the calibration process will take 

approximately 8 minutes, during which time the green LED below 
the display will flash repeatedly. Once calibration is complete the 

sensor will revert to its normal display mode.

14.

Once the display has returned to normal, turn the gas off and 
disconnect the tubing connection to complete the process.

Conversion Factors

Dew Point and Relative Humidity (RH)

While the measure of relative humidity is highly dependent on the 

temperature of air, the measurement of dew point is not. We can use dew 
point to predict what the relative humidity of air will be at a particular 
temperature. For a designer or controls contractor, this means that dew 

point can be used to predict and control water concentrations in air to 
ensure a target relative humidity is not exceeded for a design indoor 
temperature condition. The conversion factors chart, as shown in 
Figure 3 below, can be used to establish dew point set points for various 

common target relative humidity levels over a range of typical 
conditioned space temperatures.

Figure 3:  Conversion Factors - Chart

cc

/

m

i

n

AIR

200

100

60

ON

O

F

F

5

10

15 20

25

30

clear

mode

enter

Male Luer

F

itting

F

lo

w

meter

16L

B

ottle

Nitrogen

T

ie

W

rap

Regulator

Gas

P

ort

12

” T

u

b

ing

5

F

oot

o

f T

u

b

ing

F

lo

w

meter

K

no

b

Regulator

K

no

b

Содержание Telaire Vaporstat 9002

Страница 1: ...he mount ing plate as follows 1 Prepare for installation by using the mounting holes configured for US or European junction boxes 2 Use the mounting plate as a template to mark the mounting holes 3 Secure the mounting plate to the wall or junction box and make the necessary wire connections 4 Mount the Controller on the base by aligning the top clips and then securing to the bottom clips A snap so...

Страница 2: ...and remain at the initial factory settings Factory Settings Adjustments Required for Installation Select Display Units F dew point or grains lb Factory set for dew point The units selected for display on the LCD will also correspond to the units used for the analog output and relay setting of the Model 9002 Set Elevation Since the density of air and gases are affected by altitude a correction must...

Страница 3: ...tput Scaling Units The top line of the display will read OUTPUT 1 Use the Arrow button to select if you want to scale the measurement range to V or mA It should be noted that the scaling adjustment you select will be applied to both the voltage and mA output For example if you program 0 5 volt output for a 0 to 80 F measurement range the mA output will be 4 12 mA for a 0 to 80 F measurement range ...

Страница 4: ...ly calibrated The calibration kit is not necessary to perform a Single Point Calibration Zero Point Calibration This calibration mode allows the user to calibrate the sensor to a verified zero concentration of water vapor The display may not indicate 0 0 during the calibration process which is normal Calibration Procedure NOTICE Use of cellular telephones or radio transceivers within two 2 feet of...

Страница 5: ...r 5 seconds The word CAL will appear in the top line of the LCD display 11 Use the Arrow buttons to toggle to the ZERO calibration mode 12 When ZERO is displayed press the enter button to initiate the calibration process 13 Once the Enter button is pressed the calibration process will take approximately 8 minutes during which time the green LED below the display will flash repeatedly Once calibrat...

Страница 6: ...utput Normally Open and Normally Closed SPDT Gold Bifurcated 2A max 24 V Limited Warranty 18 months see warranty card for details Sensor Rated Life 15 years Accessories 2076 Calibration Kit Warranty Repairs GE Sensing will repair Telaire product that fails to meet the terms provided for in the Return and Warranty Policy Statement See http www gesensing com service brochures htm Warranty period sha...

Отзывы: