GE Sievers 900 Скачать руководство пользователя страница 2

Sievers 900 and 5310 C Series On-Line TOC Analyzer Installation Instructions

To avoid exposure to the chemical reagents, 

wear 

acid-resistant gloves, protective clothing, and safety 
goggles or a face shield when changing the reagent 
supplies.

Reagent containers are for single-use only – 
DO NOT REFILL.

 Refilling or reusing reagent containers 

will void all Analyzer and parts warranties and nullify 
any performance claims.

Operation of the Analyzer without the in-line filter 

on the sample inlet line will damage the Analyzer

 

and void the warranty. To avoid damaging the 

Analyzer, install the filter and replace the filter 

element as needed.

See the separate “Warnings” document for  
additional important information.

Step 1

Unpack and inspect the Analyzer.

Step 2

Complete the identification records.

Step 3

Select a location for the Analyzer. Refer to the System 

Specifications chart.

WARNINGS

To avoid false TOC readings and possible damage

 to 

the Analyzer, always make sure the sample is flowing 

through the iOS System and the DI water reservoir is 

filled before starting analysis.

Figure 1 - Reagent cartridges

NOTE: 

Hazardous reagents (ammonium persulfate 

and phosphoric acid) are used in the Analyzer. 
Before installing the reagents, please read the 
Material Safety Data Sheets (MSDS) contained in 
a pouch on the top of the reagent shipping box 
for proper handling precautions and spill or leak 
procedures.

Step 5

Fill the DI water reservoir.

1. Remove the rubber inlet cover on the DI water reservoir.

2. Slide the bottle nozzle into the inlet hole, and gently squeeze the 

    bottle. Be sure to fill the reservoir until the water reaches the fill 

    line, just below the inlet.

Step 4

Install the reagent cartridges. (See Figure 1.)

The oxidizer and acid reagents are shipped from GE Analytical 

Instruments in specific packaging for safety in transit. Carefully 

read the attached MSDS sheets prior to opening the packaging. 
Within the packaging are two reagent cartridges. Each reagent 

cartridge is identified with labels that indicate which cartridge 

contains ammonium persulfate and which cartridge contains 
phosphoric acid.

1. Locate the acid supply line, indicated by the 

Acid

 label. Attach 

    the PEEK nut to the acid reagent container and tighten finger-

    tight.

2. Slide the valve on the acid container to the open position by 
    pushing the green button all the way in. Refer to the labeling on 
    the container for proper positioning.

Step 6

Install accessories and cables.

If you will be using a 900 ICR unit with your configuration, make 

sure it is installed. Follow the complete set of operation instructions 
in the 

ICR Operation and Maintenance Manual

. (See downloading 

instructions on the back page of this 

Quick Start Guide.

)

Step 6 A

Install the 900 Inorganic Carbon Remover (ICR).

Installation of the Sievers 900/5310 C On-Line TOC Analyzer requires an 
external source of AC power connected to the enclosure using a water-
tight conduit connector. The electrical connection should be performed 

by a qualified electrician. An external switch or circuit breaker is recom

-

mended to facilitate maintenance and servicing of the Analyzer. It should 
be installed near the Analyzer and be clearly marked as the disconnecting 
device for the Analyzer. (See Figure 2.)

NOTE:

 Before installing any wiring inside the Analyzer, put on a 

            grounding strap for ESD protection.

Step 6 B

Connect to a power supply.

The Analyzer has three terminal blocks, arranged horizontally 
in the electrical enclosure inside the Analyzer; remove the 
enclosure cover by loosening the front screws. Consult Tables 
1–3 for a list of functions on each terminal block. The output 

and alarm connections should be installed by a qualified 

electrician. Refer to the 

900

 or 

5310 C On-Line Operations and 

Maintenance Manual 

for complete details.

Step 6 C

Install the analog outputs and alarms.

3. Slide the acid container into the reagent enclosure, with the 
    valve pointing down and toward the front of the reagent 
    enclosure.

4. Locate the oxidizer supply line, indicated by the 

Oxidizer

 

    label. Attach the PEEK nut to the oxidizer container and 

    tighten finger-tight.

5. Slide the valve on the oxidizer container to the open 
    position by pushing the green button all the way in. Refer 
    to the labeling on the container for proper positioning.   

6. Slide the oxidizer container into the reagent enclosure, with 
    the valve pointing down and toward the front of the reagent 
    enclosure.

Table 1: Inputs and Outputs (TB2)

10

Ground, for binary input

  9

24 V (+ output, for binary input)

  8

Start/Stop - (for binary input)

  7

Start/Stop + (for binary input)

  6

Alarm 2 (NO*) (output)

  5

Alarm 2 (NC*) (output)

  4

Alarm 2 (Common) (output)

  3

Alarm 1 (NO*) (output)

  2

Alarm 1 (NC*) (output)

  1

Alarm 1 (Common) (output)

* NC = normally closed     *NO = normally open

Pin Number 

(from left)

Input/Output

Table 2: Inputs and Outputs (TB1)

8

900 Lab:  Reserved 

5310 C Lab:  Stream 1 - (input) 0

7

900 Lab:  Reserved 

5310 C Lab:  Stream 1 + (input)

6

End of Batch (NO*) (output)

5

End of Batch (NC*) (output)

4

End of Batch (Common) (output)

3

Reserved

2

Reserved

1

Reserved

* NC = normally closed     *NO = normally open

Pin Number 

(from left)

Input/Output

Figure 2 - Wiring AC Power Conduit (Arrows indicate connection points.)

Отзывы: