Sievers 900 and 5310 C
On-Line TOC Analyzers
Quick Start Guide
IMPORTANT
For more detailed instructions, you can download manuals for the Sievers 900 On-Line TOC Analyzer, the
Sievers 5310 C On-Line TOC Analyzer
,
or the Sievers Inorganic Carbon Remover (ICR) from our Web site. Go to
www.geinstruments.com, click
Library
, then
Manuals
, and make your selection.
Sievers 900 and 5310 C On-Line TOC Analyzer Installation Instructions (continued)
System Specifications*
Linear range (900)
Linear range (5310 C)
Precision
Accuracy
Analysis time
0.03 ppb - 50 ppm TOC
4 ppb - 50 ppm TOC
<1% of RSD
±2% or ±0.5 ppb, whichever is greater
4 minutes
Sample flow rate
(nominal)
Power requirements
Fuses
Analysis mode: 0.5 mL/min
Fast Flush (between samples): 1.0 mL/min
100-240 ± 10% VAC, 50 watts, 50/60 Hz
No user-replaceable fuses; built-in circuit
breaker
Sample temperature
Ambient temperature
Sample pressure
(in iOS System)
Maximum relative
humidity
1 ºC to 95 ºC**
10 ºC to 40 ºC
Up to 250 psig**
Up to 95%, noncondensing
Maximum altitude
3,000 m (9,843 ft)
Installation/Overvoltage
category
Serial (RS-232), USB, parallel printer port,
4-20 mA output, two alarms, binary output
II
Calibration stability
Typically stable for 12 months
Chemical reagents
(prepackaged)
300 mL acid; 300 mL or 150 mL oxidizer
Safety certificates
CE, ETL listed. Conforms to UL Std. 61010-1
Certified to CSA C22.2 No. 61010-1
Pollution degree
2
Display
Color, touch-sensitive LCD
Weight
16.9 kg (37.2 Ibs)
* Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
** If the sample temperature and pressure are above 60 ºC and 100 psi, the 900 and 5310 C On-Line Analyzer models with the stainless steel iOS is required.
Outputs
Step 6 C
Install the Analog outputs and alarms
(continued).
Step 9
Flush the Analyzer
Before placing the Analyzer into normal operation, you should
perform a reagent flush to remove any gas bubbles that may
have formed in the reagent lines; you should then leave the
Analyzer powered on for 12 hours to allow the DI water reservoir
to properly complete rinse down action.
Size
62.4 cm height x 45.2 cm width x 26.4 cm
depth (24.6 in x 17.8 in x 10.4 in)
Required sample line
flow rate
30 mL - 300 mL/min (for On-Line Mode)
Normal operating
environment
Intended for indoor use only; avoid
sunlight
IP Rating
IP 45
Step 7
Connect the inlet system and waste outlet.
1. Connect the 1/4” Teflon tubing with the capsule or inline
filter to the sample inlet on the iOS System. Tighten 1/4 turn
past finger-tight with a 9/16” open-end wrench. Do not over-
tighten the nut.*
*
For raw water and heavy particulates applications, additional
pre-filtration is required. Please contact GE Analytical Instruments
if you need recommendations.
2. Connect the 3/4” OD waste line tubing to the waste outlet on
the iOS System by sliding the tubing over the barb fitting.
Table 3: Serial and 4-20 mA
Inputs and Outputs (TB3)
12
4-20 mA ( - output)
11
4-20 mA ( + output)
10
Not used
9
Not used
8
Reserved
7
Reserved
6
Reserved
5
Not used
4
Serial (In)
3
Serial (Out)
Pin Number
(from left)
Output
2
Serial (Ground)
1
Not used
Step 8
Configure basic Analyzer settings.
Before using the Analyzer, you will need to configure various basic
settings. Some of these settings will not need to be changed again,
unless you move the Analyzer or reconfigure the operational environ
-
ment. Refer to the 900 or
5310 C On-Line Operation and Maintenance
Manual
for complete details.
Step 8 B
Enable password protection.
Step 8 C
Set the clock and time zone.
Step 8 D
Name the Analyzer location (optional).
Step 8 E
Set reagent expiration dates.
Step 8 F
Set up the data history.
Step 8 G
Set up the printer (pptional).
Step 8 H
Export or print constants.
Step 8 A
Power the Analyzer on.
Step 8 I
Set up data I/O.
©2016. General Electric Company. All Rights Reserved.
DQS 96003-01 Rev A
Americas
: T 800 255 6964, 303 444 2009 | [email protected]
Europe/Middle East/Africa:
T 44 (0) 161 864 6800 | [email protected]
Asia Pacific:
www.geinstruments.com
For more detailed instructions, you can download operation and maintenance manuals for the Sievers 900 On-Line Analyzer, the Sievers
5310 C On-Line TOC Analyzer, and the 900 Inorganic Carbon Remover (ICR) from our Web site. Go to www.geinstruments.com, click
Library
,
then
Manuals
, and make your selection.
Where to Find Operation and Maintenance Manuals