GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A005 - Bucky Support Installation
Page no. 409
Chapter 6
JC-PHY-A-005.fm
Job Card PHY A005 - Bucky Support Installation
Chapter 6
1
SUPPLIES
Bucky Support Assembly
2
TOOLS
3 mm allen wrench
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 15 minutes
4
SAFETY PRECAUTIONS
None
5
PRE-REQUISITES
None
6
PROCEDURE
1. Put in place the Bucky Support (1) and secure it with two screws CHC 4 x 12 on the top (2) and two
screws FHC M4 x 16 on the bottom (3) (3 mm allen wrench).
2. Install the four screw covers to the fixing points of the Bucky Support (8 mm diameter on the top,
14 mm diameter on the bottom).
1
2
3