![GE RHM 30 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/ge/rhm-30/rhm-30_specifications_128439006.webp)
6
Performance RHM 30
Max Flow Rate Qmax = 750 kg/min (1653 lb/min) and Qnom (*) = 600 kg/min (1323 lb/min)
Standard Models
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
750
1653
0.20
500
1102
0.20
100
220
0.20
35
77.2
0.20
15
33.1
0.50
Uncertainty % DP (bar)
Flow Rate kg/min
Goldline Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
600
1323
0.15
300
661
0.15
150
331
0.15
100
220
0.15
60
132
0.15
Low Flow Models (**) - selected sensors
Flow Rate
Uncertainty
kg/min
lb/min
in % of reading
750
1653
0.20
500
1102
0.20
35.0
77.2
0.20
15.0
33.1
0.50
10.0
22.0
0.60
Uncertainty % DP (bar)
Flow Rate kg/min
Uncertainty % DP (bar)
Repeatability
Better ± 0.1% of rate,
Goldline 0.05%
Density
Down to 0.5% uncertainty
Temperature
Better ± 1°C
(*) Nominal flow Qnom refers to roughly 8 - 10 m/s (26 -
33 ft/s) velocity in measuring loops for best performance.
(**) Selected sensors are only available in combination
with temperature options T1, TA, standard material and
pressure range.
• Serial/single path versions offer the same accuracy at
half the flow (Qmax serial version = 375 kg/min).
• Uncertainty of reading (incl. zero drift) indications refer
to reference conditions H2O, 18-24°C (66 - 76°F), 1 - 3
bar (15 – 45 psi) and installation according to field
manual.
• Pressure drop indications refer to H2O, with parallel
measuring loops type P1 and standard manifold
construction type.
• For calibration to customer range and / or with
improved uncertainty, please consult your local
agent..