background image

Performance RHM 15

   

 

Standard Models

Flow Rate

Uncertainty

kg/min

lb/min

in % of reading

200

441

0.20

100

220

0.20

50

110

0.20

10

22

0.20

4

8.8

0.50

Goldline Models (**) - selected sensors

Flow Rate

Uncertainty

kg/min

lb/min

in % of reading

150

331

0.10

75

165

0.10

50

110

0.10

15

33.1

0.10

7.5

16.5

0.12

Low Flow Models (**) - selected sensors

Flow Rate

Uncertainty

kg/min

lb/min

in % of reading

80

176

0.20

50

110

0.20

15

33

0.20

4

8.8

0.20

2

4.4

0.60

Repeatability  

Better ± 0.1% of rate,  

 

Goldline 0.05%   

Density  

Down to 0.5% uncertainty 

 

Temperature  

Better ± 1°C 

 

(*) Nominal flow Qnom refers to roughly 8 - 10 

m/s (26 - 33 ft/s) velocity in measuring loops 

for best performance. 

 

 

(**) Selected sensors are only available in 

combination with temperature options T1, TA, 

standard material and pressure range. 

•  Serial/single path versions offer the same accuracy at 

half the flow (Qmax serial version = 100 kg/min). 

•  No relevant pressure effect due to torsional oscillation 

and semi-circle (non-deforming) measurement 

section. 

•  Uncertainty of reading (incl. zero drift) indications refer 

to reference conditions H

2

O, 18-24°C (66 - 76°F), 1 - 3 

bar (15 – 45 psi) and installation according to field 

manual. 

•  Pressure drop indications refer to H

2

O, with parallel 

measuring loops type P1 and standard manifold block 

connections.

•  For calibration to customer range and / or with 

improved uncertainty, please consult factory. 

Uncertainty %                                                                            DP (bar)

Uncertainty %                                                                            DP (bar)

Uncertainty %                                                                            DP (bar)

Flow Rate kg/min

Flow Rate kg/min

Flow Rate kg/min

Max Flow Rate Qmax = 200 kg/min (441 lb/min), Qnom (*) = 150 kg/min (331 lb/min) 

6

Содержание RHM 15

Страница 1: ...tainty down to 0 5 Repeatability better than 0 05 Unique torsion oscillator Typical measuring ranges from 2 to 200 kg min Minimal flows as low as 1 kg min Optimized solutions for batching operation Customization possible Hazardous Area Approvals ATEX CSA Medium Sized Coriolis Mass Flowmeter Advantages Torsion oscillator design assures most stable and basically drift free measurement and increased ...

Страница 2: ... SM0 PH0 parallel serial manifold construction Type removable manifold with PTFE seals and thread connection Type removable manifold with PTFE seals and flange connection only with PM0 SM0 manifold A1 285 mm 11 22 in B1 50 mm 1 97 in Terminal box without cable gland 125 x 80 x 58 mm 4 92 x 3 15 x 2 28 in A2 300 mm 11 81 in B2 70 mm 2 76 in W 0 mm for Temperature Range T1 and TA H1 454 mm 17 87 in ...

Страница 3: ... H8 Grayloc Hub 2 GR14 400 15 75 H3 Flange JIS RF 10k 40A 1 1 2 400 15 75 J1 Flange JIS RF 20k 40A 1 1 2 400 15 75 J2 Type parallel welded measuring loops without seals and flange connection PFT parallel sealless construction with thread connection Process Connection Face to face length L Order Code mm in Standard Female Thread G 3 4 400 15 75 G1 Female Thread 3 4 NPT 400 15 75 N1 Swagelok Tube in...

Страница 4: ...material B Ni82CrSiBFe 970 1000 which are not as corrosion resistant as the piping material 1 4571 316Ti Fully welded joints for corrosive liquids and higher pressure ratings can be provided please consult your local agent Weight in standard manifold construction with female threads approx 14 kg 31 lb Weight in standard sealless construction and 150 flanges approx 16 kg 35 lb Shipping on pallet ap...

Страница 5: ... of P2 measuring loops M1 standard material 1 4571 316Ti OD x WT 15 x 2 mm 0 59 x 0 079 in bar C psi F 352 50 5105 122 300 120 4351 248 250 210 3626 410 210 350 3046 662 pmax of P3 measuring loops M1 standard material 1 4571 316Ti OD x WT 15 x 3 mm 0 59 x 0 118 in bar C psi F 540 50 7832 122 450 120 6527 248 350 210 5076 410 300 350 4351 662 pmax of P1 measuring loops M3 optional material 2 4602 A...

Страница 6: ... are only available in combination with temperature options T1 TA standard material and pressure range Serial single path versions offer the same accuracy at half the flow Qmax serial version 100 kg min No relevant pressure effect due to torsional oscillation and semi circle non deforming measurement section Uncertainty of reading incl zero drift indications refer to reference conditions H2O 18 24...

Страница 7: ...M20 x 1 5 1 2 NPT or 3 4 NPT Max cable length between RHM and RHE is 100m 330 ft 200m 660 ft only with factory approval Material of Wetted Parts 1 4571 SS 316 Ti UNS S31635 standard 2 4602 Alloy C22 UNS N06022 Tantalum Others on request Sensor Enclosure Housing Stainless Steel 1 4301 SS 304 optional in 1 4571 SS 316Ti Others on request Protection Class IP 65 Optional IP 66 NEMA 4x Approvals ATEX E...

Страница 8: ...571 316Ti standard M3 Measuring Loops and Connection Part 2 4602 Alloy C22 Seal Less Construction only M4 Measuring Loops and Connection Part Tantalum PF0 Construction Type only Process Connection D1 Flange DIN DN25 PN40 Form C EN 1092 1 Form B1 D2 Flange DIN DN25 PN100 Form E EN 1092 1 Form B2 C1 Flange DIN DN40 PN40 Form C EN 1092 1 Form B1 C2 Flange DIN DN40 PN100 Form E EN 1092 1 Form B2 A1 Fl...