Cleaning, disinfection, and care
Cleaning the alarming unit
In case there is some dried or gross soil, perform pre-cleaning by wiping with a soft,
clean non-linting cloth saturated with lukewarm tap water (temperature range 27 to
44°C, 81 to 111°F) until visible dried or gross soil is removed.
1.
Dampen a soft non-linting cloth with detergent or use ready-to-use detergent
wipes.
Prepare the detergent according to the manufacturer’s instructions.
2.
Wring excess liquid from the cloth (wet but not dripping).
3.
Wipe the exterior surface. Ensure that the entire surface is wet and make sure all
surfaces are uniformly cleaned.
●
Pay special attention to hard-to-clean areas, like grooves and crevices. Use
cotton swabs to access hard to reach areas. Do not use any sharp tools to
clean the device.
4.
Replace soiled and/or dry wipes or non-linting cloths as needed.
5.
Repeat the cleaning until the device is visibly clean.
6.
Discard disposable wipes or non-linting cloths according to your hospital
guidelines. Do not reuse wipes.
7.
Rinse the device thoroughly using a soft, clean, non-linting cloth saturated with
lukewarm tap water (temperature range 27 to 44°C, 81 to 111°F).
8.
Dry thoroughly with a soft, clean, non-linting cloth before taking it into use.
Post-cleaning inspection
Inspect devices visually after cleaning and prior to disinfection for the following:
●
Cleanliness
■
If there is any visible soil present, repeat the cleaning procedure until no visible
soil remains.
●
Damage, including, but not limited to, corrosion (rust, pitting), discoloration,
excessive scratches, flaking, cracks, and wear
●
Missing or worn device labeling or marking
Do not use devices that are not functioning properly or are damaged, or have
unrecognizable markings, or missing or worn device labeling or marking.
Disinfection
Disinfection points to note
●
Always clean before disinfecting.
●
Always dilute cleaning and disinfectant agents according to their manufacturer’s
instructions. Always consider your hospital guidelines as well.
●
Use only the permitted substances.
●
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
●
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer’s instructions.
126
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...