![GE PanaFlow HT Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/ge/panaflow-ht/panaflow-ht_user-manual_128416029.webp)
PanaFlow HT User’s Manual
19
Chapter 2. Installation
2.5.4 Wiring the Modbus/Service Port
The XMT900 flow transmitter is equipped with a Modbus communication port for either a connection to Vitality (PC
software) or to a separate control system. The port is an RS485 interface.
IMPORTANT:
The maximum cable length for RS485 is 4000 ft (1200 m).
To wire to this RS485 serial port, refer to Figure 10 on page 11 and complete the following steps:
1.
Disconnect the main power to the unit and remove the rear cover.
2.
Install the required cable clamp in the chosen conduit hole on the side of the electronics enclosure.
3.
Feed one end of the cable through the conduit hole, wire it as shown in Figure 15 below.
4.
If wiring of the unit has been completed, reinstall the wiring cover on the enclosure and tighten the set screw.
Note:
Prior to use, the serial port must be programmed.
WARNING! Make sure all covers, with their O-Ring seals, are installed and the set screws tightened before
applying power in a hazardous environment.
Figure 15: Modbus Connections
Terminal Block -Modbus Connections
PIN Label
Description
1 MOD- Modbus/Service Port Negative
2 MOD+ Modbus/Service Port Positive
Содержание PanaFlow HT
Страница 1: ...GE Measurement Control Flow 910 294U Rev A September 2012 PanaFlow HT User s Manual ...
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Chapter 1 Introduction 4 PanaFlow HT User s Manual no content intended for this page ...
Страница 84: ...Chapter 3 Initial Setup and Programming 74 PanaFlow HT User s Manual no content intended for this page ...
Страница 98: ...88 PanaFlow HT User s Manual no content intended for this page ...
Страница 99: ...PanaFlow HT User s Manual 89 Appendix B Menu Maps Appendix B Menu Maps ...
Страница 100: ...Appendix B Menu Maps 90 PanaFlow HT User s Manual ...
Страница 137: ...128 PanaFlow HT User s Manual no content intended for this page ...
Страница 142: ...PanaFlow HT User s Manual 133 D 5 Review Menu Figure 39 Review Menu ...
Страница 143: ...134 PanaFlow HT User s Manual no content intended for this page ...
Страница 147: ...138 PanaFlow HT User s Manual no content intended for this page ...
Страница 153: ...144 PanaFlow HT User s Manual no content intended for this page ...
Страница 155: ...146 PanaFlow HT User s Manual no content intended for this page ...
Страница 161: ...Warranty 152 PanaFlow HT Ultrasonic Liquid Flowmeter User s Manual no content intended for this page ...
Страница 162: ......