PACE Series Calibration Manual
[EN] English
5
K0450 Revision B
7
Calibration Adjustment
To adjust a calibration range of the calibrator, proceed as follows:
1
Connect the calibrator for the range to be adjusted, as detailed in Calibration Check.
Note:
The calibration adjustments may be carried out in any order.
2
Select
Measured Pressure/Global Set-up/Calibration,
enter the Calibration
PIN
(4321)
.
3
Select
Sensor Correction
.
4
Select the pressure range to be corrected.
5
Select
New Calibration.
6
The display shows the first value to be set on the pressure standard and to press
OK
when the applied pressure is stable. Use the numeric keys to enter the precise applied
pressure.
Note:
The display also shows throughout this procedure the message Calibrating and the
selected pressure range.
7
Select
Accept
to store the first value and the display goes to the next pressure value to
be set.
8
Select
Repeat
to re-apply the same pressure and
Quit Calibration
to exit the
calibration of this pressure range.
9
Repeat steps (5) to (7) for the next value.
10
Carry out a calibration check to verify this procedure.
11
After completing the calibration procedures, adjust the calibration standard to
atmospheric pressure. Disconnect calibration standard from the PACE.
Содержание PACE5000
Страница 15: ...PACE Series Calibration Manual EN English 7 K0450 Revision B A 2 PACE Controller Screens...
Страница 16: ...K0450 Revision B 8 EN English A 3 PACE Indicator Menus Calibration procedure next page...
Страница 17: ...PACE Series Calibration Manual EN English 9 K0450 Revision B A 4 PACE Indicator Screens...
Страница 18: ......
Страница 19: ......
Страница 20: ...www gemeasurement com...