Section 11.0 - Drill the Anchor Holes
11.6
Lite Table Redundant Anchor Holes
The table requires a minimum of four (4) anchors, one (1) in each corner of table base.
Note:
If any obstruction is found under the proper anchoring position, stop drilling and use the alternate
position (see
). Use the alternate positions only when the case its corresponding proper
position is not possible.
Figure 1-55 Lite Table Anchor Positions (Proper and Alternative)
MOUNTING REQUIREMENTS ANCHOR P/N 2106573-3
Minimum Floor Thickness:
110 mm (4-1/3 in.)
Recommended Drilling Depth:
100 mm (4 in.)
Average Anchor Embedment:
90 mm (3-½ in.)
Minimum Anchor Embedment:
80 mm (3 in.)
Available Alternate Anchor Locations:
Yes
Shipped Anchor Size:
140 mm (5-½ in.)
Alternate Anchoring Methods:
Yes (see notes, above)
FLoor Levelness Requirement:
6 mm (1/4 in.) over 3 m (10 ft.)
Table 1-5 Lite Table Mounting Requirements
PROPER ANCHOR POSITION
ALTERNATIVE ANCHOR POSITION
TABLE
TABLE
DRIVER
TO GANTRY