Chapter 1 - Position Subsystems
Page 65
sys
tems
To adjust two leveling pads at rear end of table, please insert the screw driver into the hole located
on the bolt then rotate the bolt, refer to the
Figure 1-41 Adjust Leveling Pads
10.3.2
Cradle Center Procedure
1.) Remove a bottom cover under a cradle handle. See
Figure 1-42 Remove Bottom Cover