January 2006
3-6
Operation
Scenario 2: Measurement
Requiring the Process
Pressure Feature
Process pressure is 500 psi, above the measurement range of a typical
GE Infrastructure Sensing chilled mirror sensor. A flow meter is
installed upstream of the sensor cavity to expand the gas to be within
the measurable range of the dew point sensor. Since we wish to
measure the dew point of the process gas and dew point is pressure
dependent, an accurate dew point measurement at the process
pressure requires accommodating for the expansion to the sensor
cavity pressure by using the Optica’s Process Pressure feature (see
Figure 3-6 below).
Figure 3-6: Measurement Scenario 2
Measuring Vapor Pressure
Without
a Pressure Sensor
To measure vapor pressure without a pressure sensor, make the
following entries in the Pressure input menu section for this example:
•
Input: Use default
•
Units: psia
•
Default: 14.7 (the pressure at the dew point sensor)
In addition, to accommodate the pressure expansion, the following
entries are required in the Process section:
•
Status: enabled
•
Pressure: 500 (specify the process pressure with the same units
as the default pressure specified above)
Содержание Optica
Страница 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Страница 11: ...Chapter 1 ...
Страница 20: ...Chapter 2 ...
Страница 43: ...Chapter 3 ...
Страница 58: ...Chapter 4 ...
Страница 80: ...Chapter 5 ...
Страница 81: ...Programming the 4x40 Optica Introduction 5 1 Programming Technique 5 2 Programmable Functions 5 4 ...
Страница 94: ...Chapter 6 ...
Страница 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Страница 99: ...Chapter 7 ...
Страница 109: ...Appendix A ...
Страница 114: ...Appendix B ...
Страница 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 118: ...Humidity Equations and Conversion Chart B 3 January 2006 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 119: ...Appendix C ...
Страница 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing Optica Models D 3 ...
Страница 127: ...Appendix E ...
Страница 128: ...Glossary ...
Страница 130: ...Appendix F ...
Страница 133: ...Appendix G ...
Страница 134: ...Communicating with the OPTICA Using Ethernet Direct Communication G 1 Computer Communication G 1 ...